A New Error, A New Solution: Succesful Endoscopic Rescue Of A Type III Misdeployment During EUS-Guided Gastroenterostomy
Endoscopy · 01.06.2026 · 58 (S 03 S 03) · S312-S313
DOI: 10.1055/s-0046-1821179 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.06.2026. Verfasst von F Frigo und 8 weiteren Autor:innen.
Abstract (Originalsprache)
Text Lumen-apposing metal stent (LAMS) misdeployment occurs in ~10% of EUS-guided gastroenterostomies (EUS-GE). Type III misdeployments (proximal flange in peritoneum) usually require surgery.[ 1 ] [ 2 ] We report the case of a patient with malignant gastric outlet obstruction who experienced a type III misdeployment during EUS-GE. Using EUS/fluoroscopic guidance, a transgastric puncture with a 19G needle allowed guidewire advancement into the malpositioned LAMS and jejunal loop, enabling successful LAMS-in-LAMS placement. Post-procedure imaging confirmed correct lumen apposition and no leak. After resuming a solid diet the patient was discharged. To our knowledge, this is the first report of successful endoscopic rescue of a type III misdeployment, demonstrating its feasibility and potential to avoid surgery. Video http://data.process.y-congress.com/ScientificProcess/Data//106/660/1670/ce3ff6f8-a3cc-403b-9230-930e4861b714/Uploads/19978_Successful_Endoscopic%20Salvage%20of%20a%20Type%20III%20Misdeployment%20durin....mp4 Publication History Article published online: 05 June 2026 © 2026. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Zitieren
APA
Frigo, F., Elvo, B., Pignata, L., Drago, A., Laurenza, C., Luna, I. D., Soro, S., Dagani, D., & Grassia, R. (2026). A New Error, A New Solution: Succesful Endoscopic Rescue Of A Type III Misdeployment During EUS-Guided Gastroenterostomy. Endoscopy, 58(S 03 S 03), S312-S313. https://doi.org/10.1055/s-0046-1821179
Vancouver
Frigo F, Elvo B, Pignata L, Drago A, Laurenza C, Luna ID, et al. A New Error, A New Solution: Succesful Endoscopic Rescue Of A Type III Misdeployment During EUS-Guided Gastroenterostomy. Endoscopy. 2026;58(S 03 S 03):S312-S313. doi:10.1055/s-0046-1821179
BibTeX
@article{Frigo2026,
author = {Frigo, F and Elvo, B and Pignata, L and Drago, A and Laurenza, C and Luna, I Di and Soro, S and Dagani, D and Grassia, R},
title = {A New Error, A New Solution: Succesful Endoscopic Rescue Of A Type III Misdeployment During EUS-Guided Gastroenterostomy},
journal = {Endoscopy},
year = {2026},
volume = {58},
number = {S 03 S 03},
pages = {S312--S313},
doi = {10.1055/s-0046-1821179}
}
EUVIMED