Growth-Compatible Pediatric Heart Valve Prostheses: Impact of Geometric Modifications on Hydrodynamic Performance
The Thoracic and Cardiovascular Surgeon · 2026-01 · 74 (S 02) · S71-S98
DOI: 10.1055/s-0046-1817370 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Thoracic and Cardiovascular Surgeon, online erschienen am 17.02.2026. Verfasst von P. Fey und 1 weiteren Autor:innen.
Abstract (Originalsprache)
All articles of this category (opens in new window)
Background
Worldwide, an estimated 1.35 million infants are born each year with congenital heart disease. In children requiring early heart valve replacement, one operation is rarely definitive, as repeated interventions until adulthood are frequently required due to somatic growth and new pathologies. Consequently, valve prostheses that tolerate implantation variability and allow subsequent dilations to accommodate growth represent a pressing clinical need.
Methods
Eighteen balloon-expandable transcatheter valves were fabricated from cross-linked porcine pericardium using 3D-printed tools. Three valve sizes (18, 20, and 24 mm) were manufactured, with six samples per diameter. Within each diameter group, three prostheses followed the reference design and three incorporated cusp height modifications, comparing reduced with increased cusp height. Hydrodynamic performance was assessed in a pulse duplicator under ISO 5840–3:2021 pulsatile conditions. Coaptation height, measured in real time by intracardiac echocardiography (ICE), served as a surrogate for coaptation reserve. Each valve underwent sequential balloon dilations (16–24 mm) to simulate intraoperative annular variability and somatic growth. Co-primary endpoints, defined according to ISO 5840–3:2021, were RF ≤ 20% and EOA ≥ 0.7 to 1.45 cm 2 , depending on implantation diameter.
Results
All 18 bioprosthetic valves met the co-primary endpoints. Eleven geometries remained within ISO 5840–3:2021 requirements across the full dilation range. Valves with increased cusp height maintained higher coaptation reserve and superior competence during over-dilation, with the largest benefit at smaller diameters (mean RF: 8.2% vs. 14.5%). Intentional annular under expansion, resulting in increased free-edge length, did not compromise short-term hydrodynamic performance.
Conclusion
Cusp-free-edge length and height were key determinants of closing competence and coaptation quality. Designs with extended free edges preserved hydrodynamic performance across dilations, suggesting broader diameter compatibility during growth. Future valve designs should incorporate diameter variability and cusp dimensions to ensure competence and durability across expansion states. These findings highlight clinically relevant geometric parameters that may reduce the need for reinterventions and support the development of growth-compatible pediatric prostheses. Publication History Article published online: 17 February 2026 © 2026. Thieme. All rights reserved. Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Zitieren
APA
Fey, P., & Wentzel, C. (2026). Growth-Compatible Pediatric Heart Valve Prostheses: Impact of Geometric Modifications on Hydrodynamic Performance. The Thoracic and Cardiovascular Surgeon, 74(S 02), S71-S98. https://doi.org/10.1055/s-0046-1817370
Vancouver
Fey P, Wentzel C. Growth-Compatible Pediatric Heart Valve Prostheses: Impact of Geometric Modifications on Hydrodynamic Performance. The Thoracic and Cardiovascular Surgeon. 2026;74(S 02):S71-S98. doi:10.1055/s-0046-1817370
BibTeX
@article{Fey2026,
author = {Fey, P. and Wentzel, C.},
title = {Growth-Compatible Pediatric Heart Valve Prostheses: Impact of Geometric Modifications on Hydrodynamic Performance},
journal = {The Thoracic and Cardiovascular Surgeon},
year = {2026},
volume = {74},
number = {S 02},
pages = {S71--S98},
doi = {10.1055/s-0046-1817370}
}