EUVIMEDEUVIMED entdecken
Open Access

Comparative Long-Term Clinical Performance of Mechanical Aortic Valve Prostheses

Malin Granbom Koski, Natalie Glaser, Anders Franco-Cereceda, Ulrik Sartipy, Michael Dismorr

JAMA Network Open · 19.04.2024 · 7 (4) · e247525

DOI: 10.1001/jamanetworkopen.2024.7525 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JAMA Network Open, online erschienen am 19.04.2024. Verfasst von Malin Granbom Koski und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden. Hauptthemen laut MEDLINE: Heart Valve Prosthesis Implantation, Heart Valve Prosthesis. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Importance Aggregated data and long-term follow-up in national health data registers offer the opportunity to compare the performance of mechanical aortic prostheses within the same population.

Objective

To investigate the clinical performance of mechanical aortic valve prostheses. Design, Setting, and Participants This nationwide cohort study included all 5224 patients who underwent primary mechanical aortic valve replacement in Sweden between January 1, 2003, and December 31, 2018. Statistical analysis was performed between May and September 2023. Exposures Surgical aortic valve replacement with the On-X, Carbomedics, Bicarbon, Standard, Regent, Open Pivot, Masters, or Advantage valve models. Main Outcomes and Measures The primary outcome was all-cause mortality, and secondary outcomes were reintervention, heart failure, major bleeding, stroke, and embolic events. Regression standardization was used to account for baseline differences.

Results

Overall, 5224 patients (mean [SD] age, 56.8 [11.7] years; 3908 men [74.8%]) were included. Total follow-up time was 43 982 person-years (mean [SD], 8.4 [4.6] years; maximum, 17.2 years). After regression standardization, there was a significant difference in 10-year mortality between the Carbomedics model group (17%; 95% CI, 15%-18%), Regent model group (17%; 95% CI, 13%-20%), and Standard model group (17%; 95% CI, 14%-19%) compared with the Bicarbon model group (27%; 95% CI, 21%-34%). Conclusions and Relevance In this cohort study of mechanical valve surgical aortic replacement outcomes in Sweden, the rate of all-cause mortality was higher in the Bicarbon group than in the Carbomedics, Regent, and Standard model groups. These findings warrant further research on the long-term clinical performance of the Bicarbon valve.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Valve Prosthesis ImplantationHeart Valve ProsthesisMaleHumansMiddle AgedAortic ValveCohort StudiesProsthesis Design

Autor:innen und Institutionen

  • Department of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden.
  • Department of Cardiothoracic Surgery, Karolinska University Hospital, Stockholm, Sweden.
  • Department of Cardiology, Stockholm South General Hospital, Stockholm, Sweden.

Zitieren

APA

Granbom Koski, M., Glaser, N., Franco-Cereceda, A., Sartipy, U., & Dismorr, M. (2024). Comparative Long-Term Clinical Performance of Mechanical Aortic Valve Prostheses. JAMA Network Open, 7(4), e247525. https://doi.org/10.1001/jamanetworkopen.2024.7525

Vancouver

Granbom Koski M, Glaser N, Franco-Cereceda A, Sartipy U, Dismorr M. Comparative Long-Term Clinical Performance of Mechanical Aortic Valve Prostheses. JAMA Network Open. 2024;7(4):e247525. doi:10.1001/jamanetworkopen.2024.7525

BibTeX
@article{GranbomKoski2024,
  author = {Granbom Koski, Malin and Glaser, Natalie and Franco-Cereceda, Anders and Sartipy, Ulrik and Dismorr, Michael},
  title = {Comparative Long-Term Clinical Performance of Mechanical Aortic Valve Prostheses},
  journal = {JAMA Network Open},
  year = {2024},
  volume = {7},
  number = {4},
  pages = {e247525},
  doi = {10.1001/jamanetworkopen.2024.7525}
}