Mechanical Evaluation of a Threaded Interference Interlocking Mechanism for Angle-Stable Intramedullary Nailing
Veterinary and Comparative Orthopaedics and Traumatology · 01.09.2024 · 37 (05) · S111-S142
DOI: 10.1055/s-0044-1791302 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary and Comparative Orthopaedics and Traumatology, erschienen am 01.09.2024. Verfasst von J. Hanlon und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
Angle stable intramedullary nails improve bone healing, increase implant fatigue life and require fewer interlocking screw/bolts. This study evaluated fatigue performance and mechanical characteristics of a novel, partially threaded, interference interlocking mechanism for angle stable intramedullary nailing (NILN) compared to a commercially available, angle stable intramedullary nail design (I-Loc).
Methods
Single interlocking screw/bolt constructs of NILN and I-Loc implants were assembled using steel pipe segments as bone analog and placed through 50,000 cycles of simulated, physiologic axial or torsional loading. Entry torque, post-fatigue extraction torque, and 10th, 25,000th, and 50,000th cycle torsional toggle was assessed. Each construct was then loaded to failure in the same respective direction as fatigue testing. Complete constructs of each design were then assessed using a synthetic bone analog with a 50 mm central defect via nondestructive torsional and axial loading followed by axial load to failure.
Results
All constructs were angle-stable at all time points and withstood fatigue loading. Insertional torque, extraction torque: insertion torque ratio, and torsional yield load was 62%, 26%, and 363% lower, respectively, for the NILN interlocking screws. No differences in torsional peak load, torsional stiffness, axial yield load, axial stiffness or axial peak load were identified. No differences in complete construct angle stability, torsional stiffness, axial peak load, axial stiffness or axial yield load were identified. Discussion/Conclusion: The NILN generated angle stability, sustained simulated physiologic fatigue loading, and had similar mechanical characteristics to I-Loc implants. The NILN may be a suitable alternate mechanism for generating angle-stability in interlocking nails. Acknowledgments:: Funding for this project provided through generous donation by the DeBartolo family. Publication History Article published online: 16 September 2024 Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Hanlon, J., & Kim, S. E. (2024). Mechanical Evaluation of a Threaded Interference Interlocking Mechanism for Angle-Stable Intramedullary Nailing. Veterinary and Comparative Orthopaedics and Traumatology, 37(05), S111-S142. https://doi.org/10.1055/s-0044-1791302
Vancouver
Hanlon J, Kim SE. Mechanical Evaluation of a Threaded Interference Interlocking Mechanism for Angle-Stable Intramedullary Nailing. Veterinary and Comparative Orthopaedics and Traumatology. 2024;37(05):S111-S142. doi:10.1055/s-0044-1791302
BibTeX
@article{Hanlon2024,
author = {Hanlon, J. and Kim, S. E.},
title = {Mechanical Evaluation of a Threaded Interference Interlocking Mechanism for Angle-Stable Intramedullary Nailing},
journal = {Veterinary and Comparative Orthopaedics and Traumatology},
year = {2024},
volume = {37},
number = {05},
pages = {S111--S142},
doi = {10.1055/s-0044-1791302}
}
EUVIMED