Objective assessment of speech perception in CI patients via EEG mismatch negativity
Laryngo-Rhino-Otologie · 01.05.2024 · 103 (S 02 S 02) · S270
DOI: 10.1055/s-0044-1784899 ↗Kurz gefasst
Fachartikel in der Zeitschrift Laryngo-Rhino-Otologie, erschienen am 01.05.2024. Verfasst von Pattric Stauga und 6 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
Postoperative assessment of cochlear implant (CI) success often relies on word recognition score (WRS), using the subjective Freiburg Monosyllable Test (FBE). A WRS>60% is considered as the minimum threshold for successful rehabilitation. Using FBE can be challenging for certain groups and complicates the evaluation of the CI fitting processor significantly. An objective method for assessing speech perception involves mismatch negativity (MMN). Turgeon et al. (2014) suggests a correlation between speech-induced MMN and speech perception. MMN alone may not effectively differentiate speech perception among CI patients. In this study, an MMN paradigm will be used to classify the speech perception, considering additional information from the EEG.
Method
Speech stimuli were presented in an MMN paradigm via the CI processor to 22 CI patients. Standard stimulus and deviant were the syllables 'da' and 'ba'. CI performance was determined based on speech perception in the FBE: 'poor performers' (≤60%) and 'good performers' (>60%). A linear discriminant analysis was performed for classification, considering the latencies and amplitudes of the MMN, as well as the N1 and P2 of the standard stimulus. Result 'Poor performers' exhibited a reduced amplitude of the MMN, a latency delay of the N1, and increased amplitude of the P2 compared to 'good performers'. The classification demonstrated a significant discrimination rate of 76.9±6.0%, p
Zitieren
APA
Stauga, P., Schmidt, F., Zhang, L., Ehrt, K., Sara, M. V. B., Großmann, W., & Mlynski, R. (2024). Objective assessment of speech perception in CI patients via EEG mismatch negativity. Laryngo-Rhino-Otologie, 103(S 02 S 02), S270. https://doi.org/10.1055/s-0044-1784899
Vancouver
Stauga P, Schmidt F, Zhang L, Ehrt K, Sara MVB, Großmann W, et al. Objective assessment of speech perception in CI patients via EEG mismatch negativity. Laryngo-Rhino-Otologie. 2024;103(S 02 S 02):S270. doi:10.1055/s-0044-1784899
BibTeX
@article{Stauga2024,
author = {Stauga, Pattric and Schmidt, Florian and Zhang, Lichun and Ehrt, Karsten and Sara, Maria van Bonn-Ytrehus and Großmann, Wilma and Mlynski, Robert},
title = {Objective assessment of speech perception in CI patients via EEG mismatch negativity},
journal = {Laryngo-Rhino-Otologie},
year = {2024},
volume = {103},
number = {S 02 S 02},
pages = {S270},
doi = {10.1055/s-0044-1784899}
}
EUVIMED