EUVIMEDEUVIMED entdecken
Crossref

Pre-operative anatomical analysis estimating optimal electrode array in Cochlear Implant candidates based on CT or MRI images

Daniel Polterauer, Carmen Molenda, Maike Neuling, John-Martin Hempel, Joachim Müller

Laryngo-Rhino-Otologie · 01.05.2024 · 103 (S 02 S 02) · S264-S265

DOI: 10.1055/s-0044-1784879 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laryngo-Rhino-Otologie, erschienen am 01.05.2024. Verfasst von Daniel Polterauer und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

During the last years, individual cochlear implantat (CI) care has become an important topic. One aspect is the analysis of pre-operative imaging. Especially, the cochlear duct length (CDL) is of high interest for electrode length planning. Modern analysis tools offer automatic analysis of the whole inner ear where older tools relied on manual measurements of the key parameters (basal turn (A), height of the cochlea (H), and width of the basal turn (B)) for CDL estimation. Material and methods We analyzed 159 inner ears using OTOPLAN4 on CT images in clinical routine on CI candidates. Partly automatic analyzation of the inner ear failed. In these cases, an experienced OTOPLAN user performed manual analyzation on CT or if not available on MRI images up to 1.0mm slice thickness. To ensure a reliable manual analyzation we used a large monitor of high resolution, brightness, and high-quality colors (HUAWEI MateView GT 34"). We intend to analyze postoperative outcome in comparison to anatomical data, electrode array length, and preoperative conditions like etiology or residual hearing.

Results

The key parameters could be measured in 153 ears with A=9.16±0.57 mm, H=4.17±0.42 mm, B=6.82±0.60 mm, and CDL=36.11±2.56 mm. 3 ears were implanted with a Flex34 electrode array. 39 ears with a CDL of 35.93±2.87 mm were implanted with a FlexSoft or Standard. 16 ears with a CDL of 35.95±2.60 mm were implanted with a Flex28. 6 ears with a CDL of 36.71±1.45 mm were implanted with Cochlear electrode arrays.

Conclusion

A high variability in preoperative data like anatomical data, etiology, and residual hearing need to be considered when choosing the optimal electrode array. Therefore, a much larger pool of patients is needed to verify what influences CI speech understanding. Publication History Article published online: 19 April 2024 © 2024. Thieme. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Polterauer, D., Molenda, C., Neuling, M., Hempel, J., & Müller, J. (2024). Pre-operative anatomical analysis estimating optimal electrode array in Cochlear Implant candidates based on CT or MRI images. Laryngo-Rhino-Otologie, 103(S 02 S 02), S264-S265. https://doi.org/10.1055/s-0044-1784879

Vancouver

Polterauer D, Molenda C, Neuling M, Hempel J, Müller J. Pre-operative anatomical analysis estimating optimal electrode array in Cochlear Implant candidates based on CT or MRI images. Laryngo-Rhino-Otologie. 2024;103(S 02 S 02):S264-S265. doi:10.1055/s-0044-1784879

BibTeX
@article{Polterauer2024,
  author = {Polterauer, Daniel and Molenda, Carmen and Neuling, Maike and Hempel, John-Martin and Müller, Joachim},
  title = {Pre-operative anatomical analysis estimating optimal electrode array in Cochlear Implant candidates based on CT or MRI images},
  journal = {Laryngo-Rhino-Otologie},
  year = {2024},
  volume = {103},
  number = {S 02 S 02},
  pages = {S264--S265},
  doi = {10.1055/s-0044-1784879}
}
Kostenlos starten →