EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte Studie

New automated drying method reduces residual fluid in endoscope channels and operational time in a monocentric randomized controlled trial

F. Michael, M. Jung, M. Reimers, C. Oschwald, M. Friedrich-Rust

Endoscopy · 01.04.2024 · 56 (S 02 S 02) · S85

DOI: 10.1055/s-0044-1782873 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Endoscopy, erschienen am 01.04.2024. Verfasst von F. Michael und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

International guidelines currently allow different ways of drying endoscopes in the context of reprocessing in the absence of sufficient literature. The aim of the present trial was to compare a new dedicated drying device (PlasmaTyphoon+, PT) with the drying programme of a washer-disinfector for endoscopes (EWD).

Methods

Endoscopes were randomised into two groups after patient use and guideline-compliant reprocessing, specifically to evaluate the drying step of the complex reprocessing, drying and storage process. Duodenoscopes, gastroscopes, colonoscopes, bronchoscopes, EBUS and EUS devices from three companies were included. In the EWD group, reprocessing was done automatically with laminar constant compressed medical air for about 15 minutes. In the PT group, after an unavoidable drying in the EWD for two and a half minutes, drying was done using laminar and turbulent airflow for an additional two and a half minutes. After automatic drying, endoscope channels were dried manually via an adapter to determine the weight of the residual fluid using a laboratory scale. The endoscopes were then stored hanging for at least three days. This was followed by microbiological cultivation. Endpoints of the study were microbiological contamination of the endoscope channels, residual fluid after drying and operational time.

Results

230 endoscopes were included in this interim evaluation. After exclusion of one colonoscope due to early reuse, 229 endoscopes were included in the evaluation (PT: 219; EWD: . 210). PT (29, 24%) and EWD (26, 23%) had no significant difference in microbiological contamination (p=0.9; OR: 1.04 (95%-CI: 0.6-1.9). All bacteria detected were skin and environmental low-concerning bacteria and below a clinically relevant threshold of 20 colony forming units. Both residual fluid in endoscope channels (PT: 5 (4%); 0.2 g±0.1 g vs EWD: 92 (86%); 1.1 g±1.2 g) and operational time (PT: 05:57 min±17 sec vs EWD; 16:06 min±25 sec) were significantly different favoring PT (all p

Zitieren

APA

Michael, F., Jung, M., Reimers, M., Oschwald, C., & Friedrich-Rust, M. (2024). New automated drying method reduces residual fluid in endoscope channels and operational time in a monocentric randomized controlled trial. Endoscopy, 56(S 02 S 02), S85. https://doi.org/10.1055/s-0044-1782873

Vancouver

Michael F, Jung M, Reimers M, Oschwald C, Friedrich-Rust M. New automated drying method reduces residual fluid in endoscope channels and operational time in a monocentric randomized controlled trial. Endoscopy. 2024;56(S 02 S 02):S85. doi:10.1055/s-0044-1782873

BibTeX
@article{Michael2024,
  author = {Michael, F. and Jung, M. and Reimers, M. and Oschwald, C. and Friedrich-Rust, M.},
  title = {New automated drying method reduces residual fluid in endoscope channels and operational time in a monocentric randomized controlled trial},
  journal = {Endoscopy},
  year = {2024},
  volume = {56},
  number = {S 02 S 02},
  pages = {S85},
  doi = {10.1055/s-0044-1782873}
}
Kostenlos starten →