Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System
World Journal of Nuclear Medicine · 2024-03 · 23 (01) · 010-016
DOI: 10.1055/s-0044-1778709 ↗Kurz gefasst
Fachartikel in der Zeitschrift World Journal of Nuclear Medicine, online erschienen am 22.01.2024. Verfasst von Manoj Kumar Singh und 3 weiteren Autor:innen. Beteiligte Einrichtung: Medikabazaar, Technopolis Knowledge Park, Mumbai, Maharashtra, India. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aim
The objective of this study includes the NEMA (National Electrical Manufacturer Association) NU2–2018 performance evaluation of the uMIvista PET-CT (positron emission tomography-computed tomography) system.
Methods
The latest NEMA NU2–2018 guidelines have been followed for the evaluation of performance parameters of this PET-CT scanner: axial, tangential, and radial spatial resolution, sensitivity, counting losses, scatter, randomness, random and counting loss correction, image quality, time and energy resolution, image uniformity, and image registration alignment post installation of country first uMIvista PET-CT.
Results
The measured NEMA sensitivity of the uMIvista PET scanner was 12.053 cps/kBq. The spatial resolutions of the PET were measured as tangential, radial, and transaxial spatial resolutions at 10 mm, with 3.01 mm, 2.95 mm, and 2.93 mm, respectively; at 100 mm, with 3.17 mm, 3.42 mm, and 3.05 mm, respectively; and at 200 mm, with 3.65 mm, 4.54 mm, and 3.17 mm, respectively, at full-width half-maximum (FWHM); while at full-width tenths-maximum (FWTM), the values at 10 mm were 5.79 mm, 5.57 mm, and 5.69 mm, respectively, and at 100 mm were 5.59 mm, 5.96 mm, and 5.91 mm, respectively. The measured time-of-flight (TOF) timing resolution was 302.294 ps and the measured energy resolution was 11.76% with FWHM and FWTM.
Conclusion
The NEMA NU2–2018 performances of this TOF-integrated digital PET-CT system are extremely good in all parameters.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medikabazaar, Technopolis Knowledge Park, Mumbai, Maharashtra, India.
- Gujarat Imaging Centre, Ahmedabad, Gujarat, India.
ORCID-Profile: Manoj Kumar Singh
Zitieren
APA
Singh, M. K., Agarwal, K. K., Vishwakarma, M., & Patel, H. (2024). Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System. World Journal of Nuclear Medicine, 23(01), 010-016. https://doi.org/10.1055/s-0044-1778709
Vancouver
Singh MK, Agarwal KK, Vishwakarma M, Patel H. Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System. World Journal of Nuclear Medicine. 2024;23(01):010-016. doi:10.1055/s-0044-1778709
BibTeX
@article{Singh2024,
author = {Singh, Manoj Kumar and Agarwal, Krishan Kant and Vishwakarma, Manish and Patel, Hemant},
title = {Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System},
journal = {World Journal of Nuclear Medicine},
year = {2024},
volume = {23},
number = {01},
pages = {010--016},
doi = {10.1055/s-0044-1778709}
}
EUVIMED