Introduction

The String of Pearls (SOP) is a locking plate system used in veterinary orthopaedic surgery. Bending tees are used to prevent screw hole deformation during SOP contouring. Some surgeons keep the bending tees in empty nodes over nonload sharing fractures to theoretically improve construct stability. We hypothesized that the bending stiffness in a plate-bone construct would be similar or higher with bending tees kept over the gap versus empty nodes. Materials and Methods : Delrin tubes, 12-hole SOP plates, and 4 bicortical screws on each tube segment were assembled in a jig to create a simulated femoral fracture gap model. Four-point bending was performed with an Instron Universal Testing Machine (UTM). Twenty-four plate-bone constructs were tested with 12 in medial-lateral (M-L) bending (6 with tees and 6 without tees) and 12 in cranial-caudal (Cr-Ca) bending (6 with tees and 6 without tees).

Results

Bending stiffness (mean + SD) of the Cr-Ca group was 59.11 + 1.98 with tees and 59.25 + 1.69 without tees ( p = 0.88). In the M-L group, bending stiffness was 42.88 + 0.98 with tees and 40.82 + 1.05 without tees ( p = 0.0055). Discussion/Conclusion : Bending stiffness is higher in Cr-Ca bending compared to M-L bending. M-L bending stiffness was statistically stronger with bending tees over the simulated gap than without tees. There was no significant difference in Cr-Ca bending with or without tees. Future studies for cyclic bending, torsional, and axial compression tests are required to investigate the value of fitting bending tees in nonload sharing fracture repair. Acknowledgements : We acknowledge a research product discount from Orthomed USA in purchasing the implants. We acknowledge Dr. Erik Hofmeister who provided advice for methodology and statistics. Publication History Article published online: 11 September 2023 © 2023. Thieme. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Lu, P., Corriveau, K., Farag, R., Kuo, K., & Matz, B. (2023). Comparison of Bending Stiffness between String of Pearls (SOP) Plate-Bone Constructs with and without Bending Tees. Veterinary and Comparative Orthopaedics and Traumatology, 36(05), A1-A27. https://doi.org/10.1055/s-0043-1775643

Vancouver

Lu P, Corriveau K, Farag R, Kuo K, Matz B. Comparison of Bending Stiffness between String of Pearls (SOP) Plate-Bone Constructs with and without Bending Tees. Veterinary and Comparative Orthopaedics and Traumatology. 2023;36(05):A1-A27. doi:10.1055/s-0043-1775643

BibTeX
@article{Lu2023,
  author = {Lu, P.H. and Corriveau, K.M. and Farag, R. and Kuo, K. and Matz, B.},
  title = {Comparison of Bending Stiffness between String of Pearls (SOP) Plate-Bone Constructs with and without Bending Tees},
  journal = {Veterinary and Comparative Orthopaedics and Traumatology},
  year = {2023},
  volume = {36},
  number = {05},
  pages = {A1--A27},
  doi = {10.1055/s-0043-1775643}
}
Kostenlos starten →