Study on bacterial contamination and effectiveness of UV light-based reprocessing of ultrasound probes

Alexander Stefan Rudhart, Luisa Kremper, Frank Günther, Christian Sterr, Alexander Boris Stuck, Stephan Hoch

Laryngo-Rhino-Otologie · 01.05.2023 · 102 (S 02 S 02) · S211

DOI: 10.1055/s-0043-1767136 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laryngo-Rhino-Otologie, erschienen am 01.05.2023. Verfasst von Alexander Stefan Rudhart und 5 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Sonography is a basic imaging procedure in otorhinolaryngology and is frequently used in everyday clinical practice. To avoid cross-infections, adequate disinfection of the ultrasound probes is highly relevant. However, there are usually no standardized and regulated technical procedures available. Therefore, reprocessing is often inadequate. The present study should investigate the benefits of UV-C light-based reprocessing of linear ultrasound probes. Material and methods Bacterial contamination on the relevant parts of the ultrasound probe was analyzed after patients’ treatment, before and after reprocessing with UV-C light. Microbiological examinations were done 20 times in each case. The reprocessing process consisted of pre-cleaning with a dry wipe followed by UV-C light disinfection for 45 seconds with a prototype device (UV Smart Technologies, NL).

Results

After use on the patient, all ultrasound probes showed a bacterial contamination, with an average of 68.1±56.4 colony-forming units (CFU). Six additional plates were completely bacterially overgrown (>300 CFU) and therefore no longer available for a quantitative analysis. After UV-C-based reprocessing, an average contamination of 2.6±3.3 CFU was found on the ultrasound probes, while 8 of 20 samples showed 0 CFU. On the ultrasonic probes with residual contamination, only bacteria of the skin flora were detectable after disinfection.

Conclusion

The present work reveals a clinically relevant bacterial contamination on the surface of the ultrasound probes after use on the patient. Reprocessing by UV-C light seems to be a suitable, standardized method of disinfection here. In addition, there are relevant ecological and economic advantages compared to conventional methods. UV-Smart Technologies B.V. Publication History Article published online: 12 May 2023 Georg Thieme Verlag Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Rudhart, A. S., Kremper, L., Günther, F., Sterr, C., Stuck, A. B., & Hoch, S. (2023). Study on bacterial contamination and effectiveness of UV light-based reprocessing of ultrasound probes. Laryngo-Rhino-Otologie, 102(S 02 S 02), S211. https://doi.org/10.1055/s-0043-1767136

Vancouver

Rudhart AS, Kremper L, Günther F, Sterr C, Stuck AB, Hoch S. Study on bacterial contamination and effectiveness of UV light-based reprocessing of ultrasound probes. Laryngo-Rhino-Otologie. 2023;102(S 02 S 02):S211. doi:10.1055/s-0043-1767136

BibTeX
@article{Rudhart2023,
  author = {Rudhart, Alexander Stefan and Kremper, Luisa and Günther, Frank and Sterr, Christian and Stuck, Alexander Boris and Hoch, Stephan},
  title = {Study on bacterial contamination and effectiveness of UV light-based reprocessing of ultrasound probes},
  journal = {Laryngo-Rhino-Otologie},
  year = {2023},
  volume = {102},
  number = {S 02 S 02},
  pages = {S211},
  doi = {10.1055/s-0043-1767136}
}
Kostenlos starten →