Double endoscope technique for EUS-guided gastroenteroanastomosis using lumen-apposing metal stents (LAMS) in clinical practice

A. Sanchez-Yague, R. Rivera-Irigoin, A. Gonzalez Canoniga, C. Lopez Muñoz, A. Perez-Aisa

Endoscopy · 01.04.2023 · 55 (S 02 S 02) · S345

DOI: 10.1055/s-0043-1765989 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.04.2023. Verfasst von A. Sanchez-Yague und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

To review a technique using two endoscopes and instillation of water to create a EUS-guided gastroenteroanastomosis with a LAMS in a single step.

Methods

This was a clinical practice study. Informed consent was obtained for all procedures in the two centers. Either a ultrathin or a regular endoscope were advanced past the obstruction and left there, an echoendoscope was advanced to the stomach in parallel, water was instilled through the working channel of the disconnected endoscope, once an appropriate intestinal loop was targeted access was obtained with the energized tip of the stent delivery system and a 15mm LAMS was deployed. The procedure time from the insertion of the first endoscope to the release of the stent was calculated in the total series and compared in the first and last three procedures using the U Mann-Whitney test ([ Fig. 1 ]). Fig. 1 Results Fourteen procedures were identified, 12 using an ultrathin and 2 using a regular endoscope. It was possible to complete the technique in 85,7%(12/14). The overall mean time was 617,25seconds while the mean time decreased from 738seconds in the first three to 415seconds in the last three. There was a trend towards a faster procedure when experience was gained. Two procedures could not be completed due to impossibility to advance the endoscope past the obstruction. There were no immediate complications.

Conclusions

The use of an endoscope (ultrathin or regular) results in an effective and fast technique for the creation of an EUS-guided gastroenteroanastomosis especially when experience in the technique is acquired. Nevertheless, the technique still features a non-negligible rate of incomplete procedures mainly related to the obstruction degree. Publication History Article published online: 14 April 2023 © 2023. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Sanchez-Yague, A., Rivera-Irigoin, R., Canoniga, A. G., Muñoz, C. L., & Perez-Aisa, A. (2023). Double endoscope technique for EUS-guided gastroenteroanastomosis using lumen-apposing metal stents (LAMS) in clinical practice. Endoscopy, 55(S 02 S 02), S345. https://doi.org/10.1055/s-0043-1765989

Vancouver

Sanchez-Yague A, Rivera-Irigoin R, Canoniga AG, Muñoz CL, Perez-Aisa A. Double endoscope technique for EUS-guided gastroenteroanastomosis using lumen-apposing metal stents (LAMS) in clinical practice. Endoscopy. 2023;55(S 02 S 02):S345. doi:10.1055/s-0043-1765989

BibTeX
@article{SanchezYague2023,
  author = {Sanchez-Yague, A. and Rivera-Irigoin, R. and Canoniga, A. Gonzalez and Muñoz, C. Lopez and Perez-Aisa, A.},
  title = {Double endoscope technique for EUS-guided gastroenteroanastomosis using lumen-apposing metal stents (LAMS) in clinical practice},
  journal = {Endoscopy},
  year = {2023},
  volume = {55},
  number = {S 02 S 02},
  pages = {S345},
  doi = {10.1055/s-0043-1765989}
}
Kostenlos starten →