Aims
Failure to reprocess reusable endoscopes has been reported as a major threat to patient health in digestive endoscopy. Single-use duodenoscopes have been successfully used to perform ERCP although they present high costs and difficulties in disposal. For these reasons it is necessary to rationalize their use. Liver transplanted patients are immunosuppressed and they are ideal candidates for the use of these devices. We have evaluate feasibility, safety, and performance of single-use duodenoscope in this subgroup of patient.
Methods
This case series study was carried out in the Digestive Endoscopy Unit of Pisa University Hospital between July 2020-2022. We selected liver transplanted patients with positivity to KPC infection and ASGE grade for the complexity of ERCP of 3-4. EXALT Model D (Boston Scientific Corporation, USA) was used for all the procedures. Information on ERCP was collected and ompared with the average at our center using a reusable duodenoscope. The performance of the disposable duodenoscope was evaluated with a 10-point VAS score.
Results
We selected 25 patients. ASGE grade of procedural difficulty was grade 3 in 68% and grade 4 in 32% of cases. Adverse effects are comparable with reusable duodenoscopes. Comparison between the single-use duodenoscope and our center's annual average with reusable duodenoscopes for ERCP showed no significant differences. Operators expressed positive evaluations of the performance of single-use duodenoscopes with the exception of suction and insufflation performance and radiological image quality.
Conclusions
Liver transplant patients might be an ideal target for the use of single-use duodenoscope. Publication History Article published online: 14 April 2023 © 2023. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Gambaccini, D., Ruggiero, L., Casini, B., Bolognesi, V., Natali, V., Tapete, G., Bellini, M., & Marciano, E. (2023). Single-use duodenoscopes in extreme conditions: ERCP in liver transplanted patients. Endoscopy, 55(S 02 S 02), S90. https://doi.org/10.1055/s-0043-1765210
Vancouver
Gambaccini D, Ruggiero L, Casini B, Bolognesi V, Natali V, Tapete G, et al. Single-use duodenoscopes in extreme conditions: ERCP in liver transplanted patients. Endoscopy. 2023;55(S 02 S 02):S90. doi:10.1055/s-0043-1765210
BibTeX
@article{Gambaccini2023,
author = {Gambaccini, D. and Ruggiero, L. and Casini, B. and Bolognesi, V. and Natali, V. and Tapete, G. and Bellini, M. and Marciano, E.},
title = {Single-use duodenoscopes in extreme conditions: ERCP in liver transplanted patients},
journal = {Endoscopy},
year = {2023},
volume = {55},
number = {S 02 S 02},
pages = {S90},
doi = {10.1055/s-0043-1765210}
}
EUVIMED