Fachartikel in der Zeitschrift Frontiers in Pediatrics, online erschienen am 31.03.2026. Verfasst von Sapna Khemka und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology, Hepatology & Nutrition, Nationwide Children's Hospital, Columbus, Ohio, United States. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Endoscopic retrograde cholangiopancreatography (ERCP) is a key diagnostic and therapeutic tool for hepato-pancreato-biliary conditions. Although widely performed in adults, ERCP in pediatric populations is less common and poses unique challenges, particularly concerning infection control. Reusable duodenoscopes have been linked to patient-to-patient transmission of multidrug-resistant organisms (MDROs), leading to increased interest in single-use duodenoscopes. This study describes the initial experience using single-use duodenoscopes in ERCP at a single pediatric quaternary care center.
Methods
A retrospective chart review was conducted on all patients who underwent ERCP with the EXALT™ Model D single-use duodenoscope at a single pediatric center from November 1, 2022, to April 15, 2025. Demographics, procedural details, ERCP complexity scores, and post-procedure complications were collected. Descriptive statistics were used for analysis.
Results
Twenty-eight patients (mean age 16 years; 64% female) underwent ERCP with a single-use duodenoscope. Biliary cannulation was successful in 100% of cases. Most procedures were ASGE grade 1 or 2. Stone extraction was performed in 27 cases, stent removal in 25, and new stent placement in 3. Complications occurred in 4 patients (14%): abdominal pain ( n = 2), post-ERCP pancreatitis ( n = 1), and cholangitis ( n = 1), with no cases of post-procedure bleeding or MDRO infections. No procedures required conversion to reusable duodenoscopes.
Conclusions
Single-use duodenoscopes were effective and safe for ERCP in a pediatric cohort, achieving high technical success and low complication rates. While more studies are needed, particularly in younger or smaller patients and at higher ASGE grades (3 and 4), single-use duodenoscopes may offer a promising alternative to reusable devices, especially in reducing infection risk and simplifying scope maintenance. Broader adoption may depend on the development of pediatric-specific devices and cost-benefit analyses in varying clinical settings.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Gastroenterology, Hepatology & Nutrition, Nationwide Children's Hospital, Columbus, Ohio, United States.
Zitieren
APA
Khemka, S., Morris, M., Winkie, C., & Khan, M. A. (2026). Single pediatric center experience with single-use duodenoscopes. Frontiers in Pediatrics, 14. https://doi.org/10.3389/fped.2026.1721719
Vancouver
Khemka S, Morris M, Winkie C, Khan MA. Single pediatric center experience with single-use duodenoscopes. Frontiers in Pediatrics. 2026;14. doi:10.3389/fped.2026.1721719
BibTeX
@article{Khemka2026,
author = {Khemka, Sapna and Morris, Micah and Winkie, Colin and Khan, Muhammad A},
title = {Single pediatric center experience with single-use duodenoscopes},
journal = {Frontiers in Pediatrics},
year = {2026},
volume = {14},
doi = {10.3389/fped.2026.1721719}
}
EUVIMED