Using a Dedicated Extremity MRI Scanner for Depicting Anatomic Structures of Common Wrist Pathologies: A Pilot Comparison with a 3-Tesla MRI Scanner
Journal of Wrist Surgery · 01.04.2023 · 12 (02) · 147-154
DOI: 10.1055/s-0042-1744366 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Wrist Surgery, erschienen am 01.04.2023. Verfasst von Qi Yin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Plastic, Reconstructive and Hand Surgery, Amsterdam UMC, The Netherlands. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
This pilot study explored the value of a dedicated extremity magnetic resonance imaging (MRI) scanner by focusing on the qualitative depiction of important wrist structures in common wrist pathologies, the overall image quality, artifacts, and participant experience in undergoing the examinations.
Methods
Images of the right wrist of 10 healthy adult volunteers were acquired with a 0.31-Tesla (T) dedicated extremity MRI and a 3-T MRI system, using a dedicated wrist coil. Images were separately evaluated by three radiologists. Paired images were randomized and graded for the visibility of anatomical details, including the triangular fibrocartilage complex (TFCC; central disc, meniscus homolog, and ulnar attachment), carpal ligaments (scapholunate [SL] and lunotriquetral [LT] ligaments), intercarpal cartilage, and median and ulnar nerves. Mean values and standard deviations of evaluation results were calculated for each sequence. Participants' experience in undergoing the examination in both MRI scanners was explored using a questionnaire.
Results
The mean values of anatomic structures and overall image quality were significantly in favor of the 3-T MRI scanner, compared with the dedicated extremity MRI scanner. With respect to patient satisfaction in undergoing the examination, the overall trend suggested that patients were more in favor of the dedicated extremity MRI scanner.
Conclusion
For defining the real clinical value of the dedicated MRI scanner in the treatment of hand and wrist pathology, studies focused on pathologies are needed, which is certainly warranted, considering the important benefits of these devices including lower costs and higher accessibility for both patients and health care providers.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Plastic, Reconstructive and Hand Surgery, Amsterdam UMC, The Netherlands.
- Department of Radiology, St. Jansdal hospital, Harderwijk, The Netherlands.
- The Hand Clinic, Amsterdam, The Netherlands.
- Department of Radiology, Reade, Amsterdam, The Netherlands.
- Department of Radiology, Amsterdam UMC, The Netherlands.
ORCID-Profile: Qi Yin
Zitieren
APA
Yin, Q., Kichari, J. R., Alebeek, A. H. J. V., Korteweg, M. A., Teunissen, B. P., & Ritt, M. J. P. F. (2023). Using a Dedicated Extremity MRI Scanner for Depicting Anatomic Structures of Common Wrist Pathologies: A Pilot Comparison with a 3-Tesla MRI Scanner. Journal of Wrist Surgery, 12(02), 147-154. https://doi.org/10.1055/s-0042-1744366
Vancouver
Yin Q, Kichari JR, Alebeek AHJV, Korteweg MA, Teunissen BP, Ritt MJPF. Using a Dedicated Extremity MRI Scanner for Depicting Anatomic Structures of Common Wrist Pathologies: A Pilot Comparison with a 3-Tesla MRI Scanner. Journal of Wrist Surgery. 2023;12(02):147-154. doi:10.1055/s-0042-1744366
BibTeX
@article{Yin2023,
author = {Yin, Qi and Kichari, Jayant R and Alebeek, Adriana H J van and Korteweg, Mies A and Teunissen, Bernd P and Ritt, Marco J P F},
title = {Using a Dedicated Extremity MRI Scanner for Depicting Anatomic Structures of Common Wrist Pathologies: A Pilot Comparison with a 3-Tesla MRI Scanner},
journal = {Journal of Wrist Surgery},
year = {2023},
volume = {12},
number = {02},
pages = {147--154},
doi = {10.1055/s-0042-1744366}
}