EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Performance Verification of High Sensitivity Analyzer for TAT, PIC, TM, and t-PAIC

Yanhong Liu, Bo Guo, Guanghui Chen, Caixia Chen, Zhen Meng, Yan Xie, Yanru Fan, Rufei Ma, Lan Gao

TH Open · 13.12.2025 · 09 (CP)

DOI: 10.1055/a-2751-8459 ↗

Kurz gefasst

Fachartikel in der Zeitschrift TH Open, erschienen am 13.12.2025. Verfasst von Yanhong Liu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Henan Provincial People's Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan Universi…. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Projects Approved for New Business and Technology Access Audits of Henan Provincial People's Hospital in 2022, Science and Technology Research Project of Henan Provincial Department of Science and Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Thrombin–antithrombin complex (TAT), a2-plasmininhibitor–plasmin complex (PIC), thrombomodulin (TM), tissue plasminogen activator–plasminogen activator inhibitor complex (t-PAIC) has been increasingly applied in clinical practice in recent years, especially in the diagnosis and treatment of diseases associated with thrombosis and hemorrhage. However, there is no universally accepted evaluation standard for the performance verification of these four indicators currently. Therefore, we designed experiments to verify the precision, trueness, carryover, linearity, and reference intervals of these four indicators. This study is expected to provide references for subsequent research in terms of data and experimental methods. According to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP15-A2, EP06-A, and C28-A, the precision, trueness, carryover, linearity, and reference intervals were evaluated. The within-laboratory CVs of TAT, PIC, TM, and t-PAIC were 3.67, 6.51, 3.64, and 2.46% on Control L and 4.68, 4.67, 5.08, and 3.87% on Control H. The assigned value of calibrations of TAT, PIC, TM, and t-PAIC were all included in the verification intervals. The biases of the four items of Calibration 1 were −6.67, −0.90, −3.58, and −6.78% and biases on Calibration 2 were −2.70, 1.63, 2.66, and −1.16%, respectively, compared with the assigned value provided by the manufacturer. The carryover rate of each indicator was less than 1%. Within the range that meets clinical use, the best fit curves of the four indicators were linear, and the correlation coefficients of all indicators were greater than 0.99. The reference intervals provided by the manufacturer were appropriate in our laboratory. The performance of HISCL-5000 analyzer for TAT, PIC, TM, and t-PAIC analysis were acceptable and the systems were suitable for clinical analysis.

Schlagwörter

Stichwörter der Autor:innen

a2-plasmininhibitor–plasmin complexperformance verificationthrombin–antithrombin complexthrombomodulinthrombosis and hemostasis

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Henan Provincial People's Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou, Henan, China.
  • Department of Critical Care Medicine, Henan Key Laboratory for Critical Care Medicine, Zhengzhou Key Laboratory for Critical Care Medicine, Henan Provincial People's Hospital, People's Hospital of Zhengzhou University,…

Förderung

  • Projects Approved for New Business and Technology Access Audits of Henan Provincial People's Hospital in 2022
  • Science and Technology Research Project of Henan Provincial Department of Science and Technology

Zitieren

APA

Liu, Y., Guo, B., Chen, G., Chen, C., Meng, Z., Xie, Y., Fan, Y., Ma, R., & Gao, L. (2025). Performance Verification of High Sensitivity Analyzer for TAT, PIC, TM, and t-PAIC. TH Open, 09(CP). https://doi.org/10.1055/a-2751-8459

Vancouver

Liu Y, Guo B, Chen G, Chen C, Meng Z, Xie Y, et al. Performance Verification of High Sensitivity Analyzer for TAT, PIC, TM, and t-PAIC. TH Open. 2025;09(CP). doi:10.1055/a-2751-8459

BibTeX
@article{Liu2025,
  author = {Liu, Yanhong and Guo, Bo and Chen, Guanghui and Chen, Caixia and Meng, Zhen and Xie, Yan and Fan, Yanru and Ma, Rufei and Gao, Lan},
  title = {Performance Verification of High Sensitivity Analyzer for TAT, PIC, TM, and t-PAIC},
  journal = {TH Open},
  year = {2025},
  volume = {09},
  number = {CP},
  doi = {10.1055/a-2751-8459}
}
Kostenlos starten →