Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination
Endoscopy International Open · 14.03.2025 · 13 (CP)
DOI: 10.1055/a-2536-8061 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 14.03.2025. Verfasst von Koen van der Ploeg und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology and Hepatology, Erasmus MC, Rotterdam, Netherlands. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Duodenoscope contamination remains a persistent problem, exposing patients to infection risks. Automation in reprocessing may limit human error, reduce workload, and increase uniformity and traceability. However, its effectiveness should be evaluated before implementation. This study assessed the impact of implementing a novel water-based automated endoscope cleaning process on duodenoscope contamination. This before-and-after intervention study compared duodenoscope cleaning methods. From January 2022 to December 2023, conventional manual cleaning was used. From January 2024 to June 2024, the AquaTYPHOON system (AT) replaced manual cleaning. Cultures from Pentax ED34-i10T2 patient-ready duodenoscopes were collected. The main outcome was the contamination rate with microorganisms of gut or oral origin (MGO). Secondary outcomes included contamination with solely gut bacteria. Non-inferiority of the AT was tested using a generalized estimating equation with a non-inferiority margin of 5%. During the manual cleaning period, 333 duodenoscope cultures of eight duodenoscopes were collected; during the AT period, 100 cultures were collected. Pre-introduction of the AT, the contamination rate with MGO was 21.6%, which fell to 16% post-introduction (risk difference: –5.6%, upper bound 90% confidence interval [CI] 6.8%). For gut bacteria, the contamination rate decreased from 14.4% to 9% (risk difference: –5.4%, upper bound 90% CI 3.9%), indicating non-inferiority. AT reduced the contamination rate with MGO, but non-inferiority was not demonstrated. For gut bacteria, AT was non-inferior to manual cleaning. These results are promising. However, future studies should confirm these findings in larger samples and explore other advantages of using the AT in duodenoscope cleaning.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Gastroenterology and Hepatology, Erasmus MC, Rotterdam, Netherlands.
- Department of Medical Microbiology and Infectious Diseases, Erasmus MC, Rotterdam, Netherlands.
- Department of Biostatistics, Erasmus MC, Rotterdam, Netherlands.
- Quality Assurance and Regulatory Affairs office Medical Technology, Erasmus MC, Rotterdam, Netherlands.
ORCID-Profile: Koen van der Ploeg, Nicole S Erler
Zitieren
APA
van der Ploeg, K., Severin, J. A., Vos, M. C., Erler, N. S., Bulkmans, A. J. C., Bruno, M., & Mason-Slingerland, B. C. G. C. (2025). Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination. Endoscopy International Open, 13(CP). https://doi.org/10.1055/a-2536-8061
Vancouver
van der Ploeg K, Severin JA, Vos MC, Erler NS, Bulkmans AJC, Bruno M, et al. Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination. Endoscopy International Open. 2025;13(CP). doi:10.1055/a-2536-8061
BibTeX
@article{vanderPloeg2025,
author = {van der Ploeg, Koen and Severin, Juliëtte A and Vos, Margreet C and Erler, Nicole S and Bulkmans, Adriana J C and Bruno, Marco and Mason-Slingerland, Bibi C G C},
title = {Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination},
journal = {Endoscopy International Open},
year = {2025},
volume = {13},
number = {CP},
doi = {10.1055/a-2536-8061}
}
EUVIMED