EUVIMEDEUVIMED entdecken
Scoping ReviewOpen Access · CC BY-NC-ND 4.0

Handheld Point-of-Care Devices for Snakebite Coagulopathy: A Scoping Review

Michael Abouyannis iD, Amy E Marriott, Emma Stars, Dianne P Kitchen, Steve Kitchen, Tim A L Woods, Benno Kreuels, John H Amuasi, Wuelton M Monteiro, Ymkje Stienstra, Subramanian Senthilkumaran, Geoff K Isbister und 3 weitere

Thrombosis and Haemostasis · 2025-05 · 125 (05) · 405-420

DOI: 10.1055/a-2407-1400 ↗

Kurz gefasst

Scoping Review in der Zeitschrift Thrombosis and Haemostasis, online erschienen am 19.11.2024. Verfasst von Michael Abouyannis und 14 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom. Hauptthemen laut MEDLINE: Snake Bites, Point-of-Care Systems, Blood Coagulation, Disseminated Intravascular Coagulation. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 79 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Venom-induced consumption coagulopathy (VICC) is a common complication of snakebite that is associated with hypofibrinogenemia, bleeding, disability, and death. In remote tropical settings, where most snakebites occur, the 20-minute whole blood clotting test is used to diagnose VICC. Point-of-care (POC) coagulation devices could provide an accessible means of detecting VICC that is better standardized, quantifiable, and more accurate. In this scoping review, the mechanistic reasons that previously studied POC devices have failed in VICC are considered, and evidence-based recommendations are made to prioritize certain devices for clinical validation studies. Four small studies have evaluated a POC international normalized ratio (INR) device in patients with Australian Elapid, Daboia russelii, and Echis carinatus envenoming. The devices assessed in these studies either relied on a thrombin substrate endpoint, which is known to underestimate INR in patients with hypofibrinogenemia, have been recalled due to poor accuracy, or have since been discontinued. Sixteen commercially available POC devices for measuring INR, activated clotting time, activated partial thromboplastin time, fibrinogen, D-dimer, and fibrin(ogen) degradation products have been reviewed. POC INR devices that detect fibrin clot formation, as well as a novel POC device that quantifies fibrinogen were identified, which show promise for use in patients with VICC. These devices could support more accurate allocation of antivenom, reduce the time to antivenom administration, and provide improved clinical trial outcome measurement instruments. There is an urgent need for these promising POC coagulation devices to be validated in prospective clinical snakebite studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Snake BitesPoint-of-Care SystemsBlood CoagulationDisseminated Intravascular CoagulationHumansInternational Normalized RatioBlood Coagulation TestsAnimalsPoint-of-Care TestingPredictive Value of TestsEquipment DesignReproducibility of Results

Autor:innen und Institutionen

  • Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom.
  • Department of Infection Biology and Microbiomes, Institute of Infection, Veterinary and Ecological Sciences, University of Liverpool, Liverpool, United Kingdom.
  • UK National External Quality Assessment Scheme for Blood Coagulation (UK NEQAS BC), Sheffield, United Kingdom.
  • Department of Implementation Research, Bernhard Nocht Institute for Tropical Medicine, Hamburg, Germany.
  • Department of Global Health, School of Public Health, Kwame Nkrumah University of Science and Technology, Kumasi, Ghana.
  • College of Health Sciences, University of the State of Amazonas, Manaus, Brazil.

ORCID-Profile: Michael Abouyannis

Zitieren

APA

Abouyannis, M., Marriott, A. E., Stars, E., Kitchen, D. P., Kitchen, S., Woods, T. A. L., Kreuels, B., Amuasi, J. H., Monteiro, W. M., Stienstra, Y., Senthilkumaran, S., Isbister, G. K., Lalloo, D. G., Ainsworth, S., & Casewell, N. R. (2025). Handheld Point-of-Care Devices for Snakebite Coagulopathy: A Scoping Review. Thrombosis and Haemostasis, 125(05), 405-420. https://doi.org/10.1055/a-2407-1400

Vancouver

Abouyannis M, Marriott AE, Stars E, Kitchen DP, Kitchen S, Woods TAL, et al. Handheld Point-of-Care Devices for Snakebite Coagulopathy: A Scoping Review. Thrombosis and Haemostasis. 2025;125(05):405-420. doi:10.1055/a-2407-1400

BibTeX
@article{Abouyannis2025,
  author = {Abouyannis, Michael and Marriott, Amy E and Stars, Emma and Kitchen, Dianne P and Kitchen, Steve and Woods, Tim A L and Kreuels, Benno and Amuasi, John H and Monteiro, Wuelton M and Stienstra, Ymkje and Senthilkumaran, Subramanian and Isbister, Geoff K and Lalloo, David G and Ainsworth, Stuart and Casewell, Nicholas R},
  title = {Handheld Point-of-Care Devices for Snakebite Coagulopathy: A Scoping Review},
  journal = {Thrombosis and Haemostasis},
  year = {2025},
  volume = {125},
  number = {05},
  pages = {405--420},
  doi = {10.1055/a-2407-1400}
}
Kostenlos starten →