Severe bleeding associated with lumen-apposing metal stent placement for walled-off necrosis: bloody memory in WONderland
Endoscopy · 01.12.2024 · 56 (S 01 S 01) · E552-E553
DOI: 10.1055/a-2335-6707 ↗Kurz gefasst
Fallbericht in der Zeitschrift Endoscopy, erschienen am 01.12.2024. Verfasst von Shuichi Tange und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology, The University of Tokyo, Graduate School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Stents, Pancreatitis, Acute Necrotizing. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japanese Foundation for Research and Promotion of Endoscopy.
Abstract (Originalsprache)
Lumen-apposing metal stents (LAMSs) may facilitate endoscopic ultrasound (EUS)-guided treatment of walled-off necrosis (WON) as a transluminal port for necrosectomy [ 1 ] [ 2 ]. However, this modality has a risk of potentially lethal bleeding [ 3 ] [ 4 ]. In this case, we experienced dreadful bleeding when conducting EUS-guided LAMS placement with balloon dilation for direct endoscopic necrosectomy (DEN) ([ Video 1 ]). Quality: mobile 360 480 720 Download Bleeding associated with endoscopic ultrasound-guided placement of a lumen-apposing metal stent for walled-off necrosis with balloon dilation.Video 1 A 41-year-old man presented with infectious WON due to severe biliary pancreatitis, and EUS-guided treatment was scheduled. No abnormal vessels were observed along the puncture route both on contrast-enhanced computed tomography (CT) ([ Fig. 1 ]) and Doppler endosonography. We placed a LAMS 15 mm in width (Hot AXIOS; Boston Scientific Japan, Tokyo, Japan) through the electrocautery-assisted transgastric puncture. To facilitate the endoscopic passage for DEN [ 5 ], we dilated the saddle of the LAMS up to 15 mm using a balloon dilator (Giga2; Century Medical, Tokyo, Japan) ([ Fig. 2 ] a ). At the end of DEN, a decent amount of blood clots was observed in the stomach, but there was only intermittent oozing around the flange of the LAMS. Subsequent CT revealed no extravasation, but the patient developed hemorrhagic shock with cardiopulmonary arrest five hours after the procedure. Repeated CT ([ Fig. 2 ] b ) and subsequent angiography ([ Fig. 2 ] c ) delineated an extravasation from the left gastric artery branch beside the LAMS, and hemostasis was achieved via vigorous coiling of the responsible artery ([ Fig. 2 ] d ). The bleeding did not recur until WON resolution, and the LAMS was removed endoscopically without bleeding. The patient was discharged without any disability ([ Fig. 3 ]). Fig. 1 Computed tomography delineating a walled-off necrosis lesion. No abnormal vessels were observed along the possible puncture route of endoscopic ultrasound-guided puncture. Fig. 2 Severe bleeding associated with endoscopic ultrasound-guided placement of a lumen-apposing metal stent (LAMS) for walled-off necrosis (WON). a Fluoroscopy showing balloon dilation of the inner lumen of the LAMS for immediate necrosectomy. After the end of the necrosectomy, intermittent blood oozing was suspected in the area adjacent to the LAMS. b Computed tomography delineating contrast extravasation (arrow) along the lesser curvature of the stomach. c Emergent angiography delineating contrast extravasation from a branch of the left gastric artery (arrow). d Successful coiling (arrow) for hemostasis. Fig. 3 The completion of endoscopic treatment of WON without recurrence of bleeding. a Necrotic tissue within WON was removed by endoscopic necrosectomy. b Fluoroscopic image of LAMS removal. c Endoscopic image after LAMS removal without significant bleeding. d Computed tomography showed resolution of WON. In this case with nearly fatal bleeding, severe bleeding was not predictable based on preprocedural imaging, highlighting the importance of watchful monitoring and support of radiologists and surgeons in case of serious complications after LAMS placement for WON. Balloon dilation of a LAMS may increase the bleeding risk, and the feasibility of this procedure should be investigated. Endoscopy_UCTN_Code_TTT_1AS_2AJ Endoscopy E-Videos https://eref.thieme.de/e-videos E-Videos is an open access online section of the journal Endoscopy , reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high-quality video and are published with a Creative Commons CC-BY license. Endoscopy E-Videos qualify for HINARI discounts and waivers and eligibility is automatically checked during the submission process. We grant 100% waivers to articles whose corresponding authors are based in Group A countries and 50% waivers to those who are based in Group B countries as classified by Research4Life (see: https://www.research4life.org/access/eligibility/ ). This section has its own submission website at https://mc.manuscriptcentral.com/e-videos . Publication History Article published online: 25 June 2024 © 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/). Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Gastroenterology, The University of Tokyo, Graduate School of Medicine, Tokyo, Japan.
- Department of Endoscopy and Endoscopic Surgery, The University of Tokyo Hospital, Tokyo, Japan.
- Department of Radiology, The University of Tokyo, Graduate School of Medicine, Tokyo, Japan.
ORCID-Profile: Shuichi Tange, Tomotaka Saito, Naminatsu Takahara, Mitsuhiro Fujishiro
Förderung
- Japanese Foundation for Research and Promotion of Endoscopy
Zitieren
APA
Tange, S., Saito, T., Nakai, Y., Hamada, T., Watanabe, Y., Takahara, N., & Fujishiro, M. (2024). Severe bleeding associated with lumen-apposing metal stent placement for walled-off necrosis: bloody memory in WONderland. Endoscopy, 56(S 01 S 01), E552-E553. https://doi.org/10.1055/a-2335-6707
Vancouver
Tange S, Saito T, Nakai Y, Hamada T, Watanabe Y, Takahara N, et al. Severe bleeding associated with lumen-apposing metal stent placement for walled-off necrosis: bloody memory in WONderland. Endoscopy. 2024;56(S 01 S 01):E552-E553. doi:10.1055/a-2335-6707
BibTeX
@article{Tange2024,
author = {Tange, Shuichi and Saito, Tomotaka and Nakai, Yousuke and Hamada, Tsuyoshi and Watanabe, Yusuke and Takahara, Naminatsu and Fujishiro, Mitsuhiro},
title = {Severe bleeding associated with lumen-apposing metal stent placement for walled-off necrosis: bloody memory in WONderland},
journal = {Endoscopy},
year = {2024},
volume = {56},
number = {S 01 S 01},
pages = {E552--E553},
doi = {10.1055/a-2335-6707}
}