Comparison of a High-Pitch Non-ECG-Gated and a Prospective ECG-Gated Protocol for Preprocedural Computed Tomography Imaging Before TAVI/TAVR
RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · 01.02.2023 · 195 (02) · 139-147
DOI: 10.1055/a-1898-6504 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, erschienen am 01.02.2023. Verfasst von Seyd Shnayien und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Charite University Hospital Berlin, Germany. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Atrial Fibrillation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 35 Literaturangaben.
Abstract (Originalsprache)
Purpose
Preprocedural computed tomography (CT) imaging before transcatheter aortic valve implantation/replacement (TAVI/TAVR) requires high diagnostic accuracy without motion artifacts. The aim of this retrospective study is to compare the image quality of a high-pitch non-electrocardiography (ECG)-gated CT protocol used in patients with atrial tachyarrhythmias with a prospectively ECG-gated CT protocol used in patients with sinus rhythm. Materials and Methods We retrospectively included 108 patients who underwent preprocedural CT imaging before TAVI/TAVR. 52 patients with sinus rhythm were imaged using a prospectively ECG-gated protocol (Group A), and 56 patients with atrial tachyarrhythmias were imaged using the high-pitch non-ECG-gated protocol (Group B). Image quality was rated subjectively by two experienced radiologists and assessed by objective parameters including radiation dose, image noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) at the levels of the aortic root and abdominal aorta.
Results
Subjective image quality was equally good with both CT protocols, and interrater agreement was substantial in both groups but tended to be higher in Group B at the level of the aortic root (Group A: κw = 0.644, Group B: κw = 0.741). With the high-pitch non-ECG-gated CT protocol, image noise was significantly increased (p = 0.001), whereas the SNR, CNR, and radiation dose were significantly decreased (p = 0.002, p = 0.003, and p < 0.001, respectively) at the level of the aortic root compared to the prospectively ECG-gated CT protocol.
Conclusion
The high-pitch non-ECG-gated protocol yields images with similar subjective image quality compared with the prospectively ECG-gated CT protocol and allows motion-free assessment of the aortic root for accurate TAVI/TAVR planning. The high-pitch non-ECG-gated protocol may be used as an alternative for preprocedural CT imaging in patients with atrial tachyarrhythmias. Key Points: Citation Format
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Charite University Hospital Berlin, Germany.
- Berlin Institute of Health at Charité - Universitätsmedizin Berlin, Berlin, Germany.
ORCID-Profile: Seyd Shnayien, Nick Lasse Beetz, Keno Kyrill Bressem
Zitieren
APA
Shnayien, S., Beetz, N. L., Bressem, K. K., Hamm, B., & Niehues, S. M. (2023). Comparison of a High-Pitch Non-ECG-Gated and a Prospective ECG-Gated Protocol for Preprocedural Computed Tomography Imaging Before TAVI/TAVR. RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 195(02), 139-147. https://doi.org/10.1055/a-1898-6504
Vancouver
Shnayien S, Beetz NL, Bressem KK, Hamm B, Niehues SM. Comparison of a High-Pitch Non-ECG-Gated and a Prospective ECG-Gated Protocol for Preprocedural Computed Tomography Imaging Before TAVI/TAVR. RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. 2023;195(02):139-147. doi:10.1055/a-1898-6504
BibTeX
@article{Shnayien2023,
author = {Shnayien, Seyd and Beetz, Nick Lasse and Bressem, Keno Kyrill and Hamm, Bernd and Niehues, Stefan Markus},
title = {Comparison of a High-Pitch Non-ECG-Gated and a Prospective ECG-Gated Protocol for Preprocedural Computed Tomography Imaging Before TAVI/TAVR},
journal = {RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren},
year = {2023},
volume = {195},
number = {02},
pages = {139--147},
doi = {10.1055/a-1898-6504}
}
EUVIMED