Improving the accuracy of bone mineral density using a multisource CBCT
Scientific Reports · 16.02.2024 · 14 (1)
DOI: 10.1038/s41598-024-54529-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift Scientific Reports, online erschienen am 16.02.2024. Verfasst von Yuanming Hu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA. Hauptthemen laut MEDLINE: Bone Density, Spiral Cone-Beam Computed Tomography. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIDCR NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Multisource cone beam computed tomography CBCT (ms-CBCT) has been shown to overcome some of the inherent limitations of a conventional CBCT. The purpose of this study was to evaluate the accuracy of ms-CBCT for measuring the bone mineral density (BMD) of mandible and maxilla compared to the conventional CBCT. The values measured from a multi-detector CT (MDCT) were used as substitutes for the ground truth. An anthropomorphic adult skull and tissue equivalent head phantom and a homemade calibration phantom containing inserts with varying densities of calcium hydroxyapatite were imaged using the ms-CBCT, the ms-CBCT operating in the conventional single source CBCT mode, and two clinical CBCT scanners at similar imaging doses; and a clinical MDCT. The images of the anthropomorphic head phantom were reconstructed and registered, and the cortical and cancellous bones of the mandible and the maxilla were segmented. The measured CT Hounsfield Unit (HU) and Greyscale Value (GV) at multiple region-of-interests were converted to the BMD using scanner-specific calibration functions. The results from the various CBCT scanners were compared to that from the MDCT. Statistical analysis showed a significant improvement in the agreement between the ms-CBCT and MDCT compared to that between the CBCT and MDCT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Physics and Astronomy, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA.
- Department of Applied Physical Sciences, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA.
- Department of Diagnostic Sciences, Adams School of Dentistry, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA.
- Department of Radiology, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA.
- Department of Physics and Astronomy, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA. zhou@email.unc.edu.
Förderung
- NIDCR NIH HHS
- NIH HHS
- National Institutes of Health
- North Carolina Biotechnology Center
Zitieren
APA
Hu, Y., Xu, S., Li, B., Inscoe, C. R., Tyndall, D. A., Lee, Y. Z., Lu, J., & Zhou, O. (2024). Improving the accuracy of bone mineral density using a multisource CBCT. Scientific Reports, 14(1). https://doi.org/10.1038/s41598-024-54529-4
Vancouver
Hu Y, Xu S, Li B, Inscoe CR, Tyndall DA, Lee YZ, et al. Improving the accuracy of bone mineral density using a multisource CBCT. Scientific Reports. 2024;14(1). doi:10.1038/s41598-024-54529-4
BibTeX
@article{Hu2024,
author = {Hu, Yuanming and Xu, Shuang and Li, Boyuan and Inscoe, Christina R and Tyndall, Donald A and Lee, Yueh Z and Lu, Jianping and Zhou, Otto},
title = {Improving the accuracy of bone mineral density using a multisource CBCT},
journal = {Scientific Reports},
year = {2024},
volume = {14},
number = {1},
doi = {10.1038/s41598-024-54529-4}
}
EUVIMED