Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry

Elizabeth Groetsema, Adam P Huntley iD, Shane Tichy, Peter T A Reilly iD, Brian H Clowers iD

Journal of the American Society for Mass Spectrometry · 2026-01-07 · 37 (1) · 238-247

DOI: 10.1021/jasms.5c00312 ↗

At a glance

Journal article in Journal of the American Society for Mass Spectrometry, published online on 2025-12-20. Written by Elizabeth Groetsema and 4 co-authors. Institution: Department of Chemistry, Washington State University, Pullman, Washington 99164, United States. Cited 1 times so far (Crossref); 58 references. Funded by: Division of Chemistry, Agilent Technologies.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Digital quadrupoles are driven with rectangular RF waveforms and, through duty cycle control, can access higher-order Mathieu space stability zones (HZs) using comparatively low voltages. Analytically, accessing these zones remains attractive as HZs exhibit higher resolving powers ((m/z)/(Δm/z)) compared to the conventional quadrupole operation in zone (1,1). Presented here is a modification of a commercial quadrupole time-of-flight (Q-TOF) equipped with a low-voltage digital waveform driver to navigate the filtering quadrupole in the HZs. We demonstrate that a digital quadrupole operating in zones (3,1) and (3,2) achieves a narrowed isolation window for analytes up to 690 and 1543 m/z, respectively, compared to conventional operational modes. Experimental trends suggest this performance to continue to approximately 900 m/z in zone (3,1) and 2500 m/z in zone (3,2). These narrowed isolation windows were utilized to generate MS/MS data from a mixture of m-hydroxybenzoylecgonine (306 m/z) and cocaine-d3 (307 m/z), which exhibited minimal chimeric nature without the use of deconvolution software. The development of hybrid systems capable of both sine and digital operation provides a mechanism to maximize selectivity of the HZs while maintaining the benefits of traditional modes of operation. Minimizing chimeric interferences is of particular importance, and the narrow isolation widths afforded by HZs are readily accessible using tandem instruments for analytes

Subjects

Author keywords

chimeric spectradigital quadrupole mass filterquadrupole mass spectrometrytandem mass spectrometry

Authors and institutions

  • Department of Chemistry, Washington State University, Pullman, Washington 99164, United States.
  • Agilent Technologies, Santa Clara, California 95051, United States.

ORCID profiles: Adam P Huntley, Peter T A Reilly, Brian H Clowers

Funding

  • Division of Chemistry
  • Agilent Technologies

Cite

APA

Groetsema, E., Huntley, A. P., Tichy, S., Reilly, P. T. A., & Clowers, B. H. (2026). Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry. Journal of the American Society for Mass Spectrometry, 37(1), 238-247. https://doi.org/10.1021/jasms.5c00312

Vancouver

Groetsema E, Huntley AP, Tichy S, Reilly PTA, Clowers BH. Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry. Journal of the American Society for Mass Spectrometry. 2026;37(1):238-247. doi:10.1021/jasms.5c00312

BibTeX
@article{Groetsema2026,
  author = {Groetsema, Elizabeth and Huntley, Adam P and Tichy, Shane and Reilly, Peter T A and Clowers, Brian H},
  title = {Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry},
  journal = {Journal of the American Society for Mass Spectrometry},
  year = {2026},
  volume = {37},
  number = {1},
  pages = {238--247},
  doi = {10.1021/jasms.5c00312}
}
Start for free →