A Highly Efficient and Automated Magnetic Bead Extraction Method Overcomes the Matrix Effect in LC–MS/MS Analysis of Human Serum Steroid Hormones

Xiaoyi Yi iD, Xiaojing Huang, Yufeng Xiong, Yingsong Wu iD

Journal of the American Society for Mass Spectrometry · 2025

DOI: 10.1021/jasms.4c00338 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the American Society for Mass Spectrometry, 2025. Verfasst von Xiaoyi Yi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Antibody Engineering, School of Laboratory Medicine and Biotechnology, Southern Medical University, Guangzhou 510515, China. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Steroids. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 42 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Profiling of steroid hormones is incredibly valuable in clinical settings for diagnosing endocrine disorders. However, the presence of matrix effects and labor-intensive manual work in LC-MS/MS analysis has hindered its routine application. In the present study, a highly efficient and automated magnetic bead extraction method was developed to address matrix effects and quantitatively profile 15 steroid hormones in human serum. Octadecyl (C18) and N-vinylpyrrolidone divinylbenzene (HLB) modified magnetic beads were compared for enriching steroids from human serum. Following enrichment, the beads were separated using a magnetic field; the matrix was cleaned, and the steroid hormones were eluted from the beads for LC-MS/MS analysis. This entire process of enrichment, cleanup, and elution was conducted automatically, making it simple, fast, and cost-effective. The results indicated that steroid hormones could be selectively enriched from human serum in just 1 min using C18 magnetic beads. The absolute matrix effect, evaluated as the relative response between human serum matrix and methanol solution, ranged from 89.2% to 113.1% for low levels, from 82.3% to 112.0% for medium levels, and from 91.7% to 111.0% for high levels. The intrabatch coefficients of variation (CVs) and interbatch CVs were between 3.1% and 13.4% and between 3.0% and 13.7%, respectively. Recoveries were between 87.6% and 114.3% for low levels, 94.0% and 105.0% for medium levels, and 91.9% and 111.7% for high levels. The clinical application was demonstrated by profiling steroid hormones in 160 pregnant women at various gestational weeks. The results suggested that the automated magnetic bead extraction method for LC-MS/MS could effectively address matrix effects in profiling steroid hormones. To our knowledge, this is the first automated magnetic bead extraction method for LC-MS/MS profiling of steroid hormones in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass SpectrometrySteroidsHumansChromatography, LiquidFemaleAutomationReproducibility of ResultsLimit of DetectionLiquid Chromatography-Mass Spectrometry

Stichwörter der Autor:innen

automationliquid chromatography−tandem mass spectrometrymagnetic bead extractionmatrix effectsteroid hormone

Substanzen

Steroids

Autor:innen und Institutionen

  • Institute of Antibody Engineering, School of Laboratory Medicine and Biotechnology, Southern Medical University, Guangzhou 510515, China.
  • Department of Laboratory Medicine, Nanfang Hospital, Southern Medical University, Guangzhou 510515, China.

ORCID-Profile: Xiaoyi Yi, Yingsong Wu

Zitieren

APA

Yi, X., Huang, X., Xiong, Y., & Wu, Y. (2025). A Highly Efficient and Automated Magnetic Bead Extraction Method Overcomes the Matrix Effect in LC–MS/MS Analysis of Human Serum Steroid Hormones. Journal of the American Society for Mass Spectrometry. https://doi.org/10.1021/jasms.4c00338

Vancouver

Yi X, Huang X, Xiong Y, Wu Y. A Highly Efficient and Automated Magnetic Bead Extraction Method Overcomes the Matrix Effect in LC–MS/MS Analysis of Human Serum Steroid Hormones. Journal of the American Society for Mass Spectrometry. 2025. doi:10.1021/jasms.4c00338

BibTeX
@article{Yi2025,
  author = {Yi, Xiaoyi and Huang, Xiaojing and Xiong, Yufeng and Wu, Yingsong},
  title = {A Highly Efficient and Automated Magnetic Bead Extraction Method Overcomes the Matrix Effect in LC–MS/MS Analysis of Human Serum Steroid Hormones},
  journal = {Journal of the American Society for Mass Spectrometry},
  year = {2025},
  doi = {10.1021/jasms.4c00338}
}
Kostenlos starten →