EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Pitfalls in the Detection of Volatiles Associated with Heated Tobacco and e-Vapor Products When Using PTR-TOF-MS

Noel Bielik, Daniela Correia, Kelly Rodrigues Crespo, Catherine Goujon-Ginglinger, Maya I Mitova iD

Journal of the American Society for Mass Spectrometry · 2024

DOI: 10.1021/jasms.4c00062 ↗

At a glance

Journal article in Journal of the American Society for Mass Spectrometry, 2024. Written by Noel Bielik and 4 co-authors. Institution: PMI R&D, Philip Morris Products S.A., Quai Jeanrenaud 5, CH-2000 Neuchâtel, Switzerland. Major MEDLINE topics: Mass Spectrometry, Volatile Organic Compounds, Electronic Nicotine Delivery Systems, Nicotiana. Cited 5 times so far (Crossref); 42 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

We investigated the applicability of proton transfer reaction-time-of-flight mass spectrometry (PTR-TOF-MS) for quantitative analysis of mixtures comprising glycerin, acetol, glycidol, acetaldehyde, acetone, and propylene glycol. While PTR-TOF-MS offers real-time simultaneous determination, the method selectivity is limited when analyzing compounds with identical elemental compositions or when labile compounds present in the mixture produce fragments that generate overlapping ions with other matrix components. In this study, we observed significant fragmentation of glycerin, acetol, glycidol, and propylene glycol during protonation via hydronium ions (H3O+). Nevertheless, specific ions generated by glycerin (m/z 93.055) and propylene glycol (m/z 77.060) enabled their selective detection. To thoroughly investigate the selectivity of the method, various mixtures containing both isotope-labeled and unlabeled compounds were utilized. The experimental findings demonstrated that when samples contained high levels of glycerin, it was not feasible to perform time-resolved analysis in H3O+ mode for acetaldehyde, acetol, and glycidol. To overcome the observed selectivity limitations associated with the H3O+ reagent ions, alternative ionization modes were investigated. The ammonium ion mode proved appropriate for analyzing propylene glycol (m/z 94.086) and acetone (m/z 76.076) mixtures. Concerning the nitric oxide mode, specific m/z were identified for acetaldehyde (m/z 43.018), acetone (m/z 88.039), glycidol (m/z 73.028), and propylene glycol (m/z 75.044). It was concluded that considering the presence of multiple product ions and the potential influence of other compounds, it is crucial to conduct a thorough selectivity assessment when employing PTR-TOF-MS as the sole method for analyzing compounds in complex matrices of unknown composition.

Subjects

MeSH terms (MEDLINE)

Mass SpectrometryVolatile Organic CompoundsElectronic Nicotine Delivery SystemsNicotianaPropylene GlycolAcetaldehydeAcetoneGlycerolHot TemperatureEpoxy CompoundsPropanols

Author keywords

PTR-MSglycerinonline analysispropylene glycolselected reagent iontime-resolved analysis

Substances

Volatile Organic CompoundsPropylene GlycolAcetaldehydeAcetoneGlycerolglycidolEpoxy CompoundsPropanolsacetol

Authors and institutions

  • PMI R&D, Philip Morris Products S.A., Quai Jeanrenaud 5, CH-2000 Neuchâtel, Switzerland.

ORCID profiles: Maya I Mitova

Cite

APA

Bielik, N., Correia, D., Rodrigues Crespo, K., Goujon-Ginglinger, C., & Mitova, M. I. (2024). Pitfalls in the Detection of Volatiles Associated with Heated Tobacco and e-Vapor Products When Using PTR-TOF-MS. Journal of the American Society for Mass Spectrometry. https://doi.org/10.1021/jasms.4c00062

Vancouver

Bielik N, Correia D, Rodrigues Crespo K, Goujon-Ginglinger C, Mitova MI. Pitfalls in the Detection of Volatiles Associated with Heated Tobacco and e-Vapor Products When Using PTR-TOF-MS. Journal of the American Society for Mass Spectrometry. 2024. doi:10.1021/jasms.4c00062

BibTeX
@article{Bielik2024,
  author = {Bielik, Noel and Correia, Daniela and Rodrigues Crespo, Kelly and Goujon-Ginglinger, Catherine and Mitova, Maya I},
  title = {Pitfalls in the Detection of Volatiles Associated with Heated Tobacco and e-Vapor Products When Using PTR-TOF-MS},
  journal = {Journal of the American Society for Mass Spectrometry},
  year = {2024},
  doi = {10.1021/jasms.4c00062}
}
Start for free →