Effects of Nano-Electrospray Ionization Emitter Position on Unintentional In-Source Activation of Peptide and Protein Ions

Samantha O Shepherd, Austin W Green, Elizabeth S Resendiz, Kenneth R Newton, Ruwan T Kurulugama, James S Prell iD

Journal of the American Society for Mass Spectrometry · 2024

DOI: 10.1021/jasms.3c00371 ↗

At a glance

Journal article in Journal of the American Society for Mass Spectrometry, 2024. Written by Samantha O Shepherd and 5 co-authors. Institution: Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States. Major MEDLINE topics: Spectrometry, Mass, Electrospray Ionization, Peptides. Cited 17 times so far (Crossref); 66 references. The full text is freely available in PubMed Central. Funded by: NIGMS NIH HHS, Division of Chemistry, National Institute of General Medical Sciences.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Native ion mobility-mass spectrometry (IM-MS) typically introduces protein ions into the gas phase through nano-electrospray ionization (nESI). Many nESI setups have mobile stages for tuning the ion signal and extent of co-solute and salt adduction. However, tuning the position of the emitter capillary in nESI can have unintended downstream consequences for collision-induced unfolding or collision-induced dissociation (CIU/D) experiments. Here, we show that relatively small variations in the nESI emitter position can shift the midpoint (commonly called the "CID50" or "CIU50") potential of CID breakdown curves and CIU transitions by as much as 8 V on commercial instruments. A spatial "map" of the shift in CID50 for the loss of heme from holomyoglobin onto the emitter position on a Waters Synapt G2-Si mass spectrometer shows that emitter positions closer to the instrument inlet can result in significantly greater in-source activation, whereas different effects are found on an Agilent 6545XT instrument for the ions studied. A similar effect is observed for CID of the singly protonated leucine enkephalin peptide and Shiga toxin 1 subunit B homopentamer on the Waters Synapt G2-Si instrument. In-source activation effects on a Waters Synapt G2-Si are also investigated by examining the RMSD between CIU fingerprints acquired at different emitter positions and the shifts in CIU50 for structural transitions of bovine serum albumin and NIST monoclonal antibody.

Subjects

MeSH terms (MEDLINE)

Spectrometry, Mass, Electrospray IonizationPeptidesIonsSerum Albumin, Bovine

Author keywords

activationcollision-induced dissociationelectrospray ionizationnative mass spectrometry

Substances

PeptidesIonsSerum Albumin, Bovine

Authors and institutions

  • Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States.
  • Agilent Technologies, 5301 Stevens Creek Blvd, Santa Clara, California 95051, United States.
  • Materials Science Institute, University of Oregon, Eugene, Oregon 97403-1252, United States.

ORCID profiles: James S Prell

Funding

  • NIGMS NIH HHS
  • Division of Chemistry
  • National Institute of General Medical Sciences
  • Agilent Technologies

Cite

APA

Shepherd, S. O., Green, A. W., Resendiz, E. S., Newton, K. R., Kurulugama, R. T., & Prell, J. S. (2024). Effects of Nano-Electrospray Ionization Emitter Position on Unintentional In-Source Activation of Peptide and Protein Ions. Journal of the American Society for Mass Spectrometry. https://doi.org/10.1021/jasms.3c00371

Vancouver

Shepherd SO, Green AW, Resendiz ES, Newton KR, Kurulugama RT, Prell JS. Effects of Nano-Electrospray Ionization Emitter Position on Unintentional In-Source Activation of Peptide and Protein Ions. Journal of the American Society for Mass Spectrometry. 2024. doi:10.1021/jasms.3c00371

BibTeX
@article{Shepherd2024,
  author = {Shepherd, Samantha O and Green, Austin W and Resendiz, Elizabeth S and Newton, Kenneth R and Kurulugama, Ruwan T and Prell, James S},
  title = {Effects of Nano-Electrospray Ionization Emitter Position on Unintentional In-Source Activation of Peptide and Protein Ions},
  journal = {Journal of the American Society for Mass Spectrometry},
  year = {2024},
  doi = {10.1021/jasms.3c00371}
}
Start for free →