Real-Time Monitoring of Chemisorption of Antibodies onto Self-Assembled Monolayers Deposited on Gold Electrodes Using Electrochemical Impedance Spectroscopy

Soraia Oliveira, Brian V Jones, Pedro Estrela, Paulo R F Rocha, Nuno Miguel Reis iD

Langmuir · 2025

DOI: 10.1021/acs.langmuir.5c01062 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Langmuir, 2025. Verfasst von Soraia Oliveira und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical Engineering, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom. Hauptthemen laut MEDLINE: Gold, Dielectric Spectroscopy, Immunoglobulin G. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: North Atlantic Treaty Organization, European Research Executive Agency, University of Bath.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Understanding protein binding to biosensing surfaces is paramount to the design and performance of biosensing devices in fields such as point-of-care testing and bioanalytics. Here, we systematically demonstrated the use of electrical impedance spectroscopy (EIS) and equivalent circuit modeling for real-time tracking of chemisorption of IgG antibody to large-area circular gold electrodes (1.3 mm2) functionalized with a self-assembled monolayer (SAM). Using 1 μg/mL IgG and 5 mM of [Fe(CN)6]3-/4-, the measured low-frequency impedance proved sensitive to both equilibrium and kinetics of antibody binding, with a slope of ∼74 kΩ/h for the first 2 h and taking approximately 4 h to reach equilibrium in a standard 6 mm-diameter well. Changes in impedance were found to be proportional to the reciprocal of the change in capacitance up to half-to-full IgG monolayer bound to the SAM. Further experiments with a flat microchannel confirmed that the low-frequency impedance and equivalent charge-transfer resistance (Rct) depend not only on antibody diffusion but also on the surface-to-volume ratio, which can represent a major challenge previously unreported for the miniaturization of EIS in microfluidic devices. This challenge arises as it requires a higher concentration of [Fe(CN)6]3-/4-, of 50 mM or above, which was found to interfere with Rct during chemisorption at low IgG concentrations. Chemisorption of IgG to SAM was confirmed with fluorescence microscopy and FTIR. This study marks, to the best of our knowledge, the first experimental demonstration of EIS as a real-time technique for quantitation of Langmuir isotherms during chemisorption of antibodies to SAM, with the potential to improve the design of EIS-based biosensors, especially those integrated into microfluidic devices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

GoldDielectric SpectroscopyImmunoglobulin GElectrodesSurface PropertiesAdsorptionBiosensing Techniques

Substanzen

GoldImmunoglobulin G

Autor:innen und Institutionen

  • Department of Chemical Engineering, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.
  • Centre for Bioengineering & Biomedical Technologies (CBio), University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.
  • Department of Life Sciences, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.
  • Department of Electronic & Electrical Engineering, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.
  • Centre for Functional Ecology-Science for People & the Planet, Associate Laboratory TERRA, Department of Life Sciences, University of Coimbra, Coimbra 3000-456, Portugal.
  • Centre of Biological Engineering (CEB), University of Minho Campus de Gualtar, Braga 4710-057, Portugal.

ORCID-Profile: Nuno Miguel Reis

Förderung

  • North Atlantic Treaty Organization
  • European Research Executive Agency
  • University of Bath
  • H2020 European Research Council

Zitieren

APA

Oliveira, S., Jones, B. V., Estrela, P., Rocha, P. R. F., & Reis, N. M. (2025). Real-Time Monitoring of Chemisorption of Antibodies onto Self-Assembled Monolayers Deposited on Gold Electrodes Using Electrochemical Impedance Spectroscopy. Langmuir. https://doi.org/10.1021/acs.langmuir.5c01062

Vancouver

Oliveira S, Jones BV, Estrela P, Rocha PRF, Reis NM. Real-Time Monitoring of Chemisorption of Antibodies onto Self-Assembled Monolayers Deposited on Gold Electrodes Using Electrochemical Impedance Spectroscopy. Langmuir. 2025. doi:10.1021/acs.langmuir.5c01062

BibTeX
@article{Oliveira2025,
  author = {Oliveira, Soraia and Jones, Brian V and Estrela, Pedro and Rocha, Paulo R F and Reis, Nuno Miguel},
  title = {Real-Time Monitoring of Chemisorption of Antibodies onto Self-Assembled Monolayers Deposited on Gold Electrodes Using Electrochemical Impedance Spectroscopy},
  journal = {Langmuir},
  year = {2025},
  doi = {10.1021/acs.langmuir.5c01062}
}
Kostenlos starten →