EUVIMEDEUVIMED entdecken
Crossref

Low-Background Cardiac Troponin I Detection via In Situ Self-Color-Changing Lateral Flow Immunoassay Enabled by H 2 O 2 -Triggered Cerium Oxide Nanoprobes

Jianying Li, Mingyue Luo, Kehui Zhang, Honghong Rao iD, Haile Liu, Ruibin Qiang, Zhentong Zhu, Marcin Frankowski iD, Zhonghua Xue iD

Analytical Chemistry · 26.05.2026 · 98 (20) · 15104-15114

DOI: 10.1021/acs.analchem.6c00895 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 12.05.2026. Verfasst von Jianying Li und 8 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Bioelectrochemistry & Environmental Analysis of Gansu Province, College of Chemistry & Chemical Engineering, Northwest Normal Unive…. Hauptthemen laut MEDLINE: Cerium, Hydrogen Peroxide, Troponin I, Metal Nanoparticles. Die Arbeit nennt 38 Literaturangaben. Gefördert von: Ministry of Education of the People's Republic of China, Gansu Education Department, Ministry of Finance.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Conventional lateral flow immunoassay (LFIA) based on gold nanoparticles (AuNPs) often suffer from low detection sensitivity. Current signal enhancement strategies, such as metal in situ growth and enzyme catalysis, still face challenges, including poor stability of exogenous chromogenic reagents (prone to aggregation/degradation), uncontrollable enhancement processes, and strong background interference. To address these critical limitations, we developed an in situ, controllable, and background-free self-color-changing enhanced LFIA (SCCE-LFIA) utilizing novel cerium oxide nanoparticles coated with a polydopamine layer (CeO 2 @PDA NPs). The core of this method involves the H 2 O 2 -triggered oxidation of Ce 3+ to Ce 4+ on the nanoprobe surface, inducing a distinct colorimetric signal change from colorless/light brown to yellow/orange directly at the test line (T line). This intrinsic chemical-state-transition-driven signal enhancement fundamentally circumvents issues of uncontrollable enhancement processes, diffusion of chromogenic products, and nonspecific adsorption, thereby eliminating severe background interference. It strongly enables rapid and sensitive detection of cardiac troponin I (cTnI), a crucial clinical biomarker, with a low visual limit of detection (vLOD) of 0.8 ng/mL with only 30 s of coloration. This work establishes a precedent for the design of in situ, controllable, and background-free signal enhancement strategies, confirms the effectiveness of CeO 2 @PDA NPs as substitutes for traditional chromogenic substrates, and paves the way for the development of accurate and reliable point-of-care testing (POCT) in analytical chemistry and clinical diagnostics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CeriumHydrogen PeroxideTroponin IMetal NanoparticlesImmunoassayHumansIndolesColorimetryColorPolymersLimit of Detection

Substanzen

Ceriumceric oxideHydrogen PeroxideTroponin IIndolesPolymerspolydopamine

Autor:innen und Institutionen

  • Key Laboratory of Bioelectrochemistry & Environmental Analysis of Gansu Province, College of Chemistry & Chemical Engineering, Northwest Normal University, Lanzhou 730070, China.
  • School of Chemical Engineering, Lanzhou City University, Lanzhou 730070, China.
  • Faculty of Chemistry, Adam Mickiewicz University, ul. Uniwersytetu Poznańskiego 8, 61-614 Poznań, Poland.

ORCID-Profile: Honghong Rao, Marcin Frankowski, Zhonghua Xue

Förderung

  • Ministry of Education of the People's Republic of China
  • Gansu Education Department
  • Ministry of Finance
  • National Natural Science Foundation of China
  • Science and Technology Specialist Project

Zitieren

APA

Li, J., Luo, M., Zhang, K., Rao, H., Liu, H., Qiang, R., Zhu, Z., Frankowski, M., & Xue, Z. (2026). Low-Background Cardiac Troponin I Detection via In Situ Self-Color-Changing Lateral Flow Immunoassay Enabled by H 2 O 2 -Triggered Cerium Oxide Nanoprobes. Analytical Chemistry, 98(20), 15104-15114. https://doi.org/10.1021/acs.analchem.6c00895

Vancouver

Li J, Luo M, Zhang K, Rao H, Liu H, Qiang R, et al. Low-Background Cardiac Troponin I Detection via In Situ Self-Color-Changing Lateral Flow Immunoassay Enabled by H 2 O 2 -Triggered Cerium Oxide Nanoprobes. Analytical Chemistry. 2026;98(20):15104-15114. doi:10.1021/acs.analchem.6c00895

BibTeX
@article{Li2026,
  author = {Li, Jianying and Luo, Mingyue and Zhang, Kehui and Rao, Honghong and Liu, Haile and Qiang, Ruibin and Zhu, Zhentong and Frankowski, Marcin and Xue, Zhonghua},
  title = {Low-Background Cardiac Troponin I Detection via In Situ Self-Color-Changing Lateral Flow Immunoassay Enabled by H 2 O 2 -Triggered Cerium Oxide Nanoprobes},
  journal = {Analytical Chemistry},
  year = {2026},
  volume = {98},
  number = {20},
  pages = {15104--15114},
  doi = {10.1021/acs.analchem.6c00895}
}
Kostenlos starten →