Development of a High-Sensitivity Electrochemical Immunoassay Using a Fully 3D-Printed Electrocatalytic Microelectrode Probe Platform
Analytical Chemistry · 31.03.2026 · 98 (12) · 9260-9272
DOI: 10.1021/acs.analchem.5c08037 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 19.03.2026. Verfasst von Niamh Docherty und 9 weiteren Autor:innen. Beteiligte Einrichtung: University of Strathclyde, Thomas Graham Building Centre for Advanced Measurement Science and Health Translation, Pure and Applied Chemistry, 295 Cat…. Hauptthemen laut MEDLINE: Electrochemical Techniques, Printing, Three-Dimensional, Troponin I. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: British Heart Foundation, University of Strathclyde.
Abstract (Originalsprache)
Electrochemical biosensors are a promising route to point-of-care diagnostics, yet their translation is hindered by the need for electrode surface functionalization, which introduces variability, increases cost and production complexity, and limits scalability and stability. Additive manufacturing using conductive filaments for rapid fabrication of three-dimensional (3D) electrodes overcomes these limitations. This study evaluates composite filaments comprising polylactic acid (PLA), carbon black (CB), and multiwalled carbon nanotubes (MWCNTs) for fused filament fabrication (FFF) of electrochemical electrodes for indirect detection via TMB+ measurement. Both the filament composition and electrode size were directly compared to determine the most suitable electrode architecture for sensitive measurements in complex samples. As a key novelty, the electrode diameter was reduced from 1 mm to the microscale to investigate the influence of electrode size on electron transfer efficiency. To improve measurement consistency and throughput, a 3D-printed assay accessory ("The Consistent Dipper") was developed to guide electrode immersion and reduce movement during amperometric measurements. PLA/MWCNT microelectrodes exhibited increased current density and reduced background noise compared to carbon black filament electrodes. The MWCNT microelectrodes were subsequently applied to a cardiac troponin I (cTnI) electrochemical immunoassay and in undiluted human serum, a cTnI detection limit of 7.4 pg mL-1 was achieved, representing an approximately 19-fold improvement compared to PLA/MWCNT macroelectrodes (140 pg mL-1). Following optimization to reduce incubation times, a clinically relevant detection limit of 85 pg mL-1 was obtained within a total assay time of 1 h. By combining enhanced electrochemical performance with low-cost, flexible FFF-printed microelectrodes, this platform provides a scalable route to rapid immunodiagnostics. This study represents the first application of microscale 3D-printed PLA/MWCNT electrodes for clinical biomarker detection using a readily manufacturable sensor system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- University of Strathclyde, Thomas Graham Building Centre for Advanced Measurement Science and Health Translation, Pure and Applied Chemistry, 295 Cathedral St, Glasgow G1 1XL, U.K.
- School of Applied Sciences, Centre for Lifelong Health, University of Brighton, Brighton, East Sussex BN2 4GJ, U.K.
- University of Strathclyde, Biomedical Engineering, Wolfson Centre, 106 Rottenrow East, G4 0NW Glasgow, U.K.
- National Measurement Laboratory at LGC The Priestley Building, Guildford, Surrey GU2 7XY, U.K.
- University of Strathclyde, Pure and Applied Chemistry, Technology Innovation Centre, 99 George Street, Glasgow G1 1RD, U.K.
ORCID-Profile: Niamh Docherty, Chloe L Miller, Daniel Macdonald, Damion K Corrigan, Bhavik Anil Patel
Förderung
- British Heart Foundation
- University of Strathclyde
Zitieren
APA
Docherty, N., Miller, C. L., Dobrea, A., Macdonald, D., Gordon, A., Pang, S., Fu, Y., Jimenez, M., Corrigan, D. K., & Patel, B. A. (2026). Development of a High-Sensitivity Electrochemical Immunoassay Using a Fully 3D-Printed Electrocatalytic Microelectrode Probe Platform. Analytical Chemistry, 98(12), 9260-9272. https://doi.org/10.1021/acs.analchem.5c08037
Vancouver
Docherty N, Miller CL, Dobrea A, Macdonald D, Gordon A, Pang S, et al. Development of a High-Sensitivity Electrochemical Immunoassay Using a Fully 3D-Printed Electrocatalytic Microelectrode Probe Platform. Analytical Chemistry. 2026;98(12):9260-9272. doi:10.1021/acs.analchem.5c08037
BibTeX
@article{Docherty2026,
author = {Docherty, Niamh and Miller, Chloe L and Dobrea, Alexandra and Macdonald, Daniel and Gordon, Alisdair and Pang, Susan and Fu, Ying and Jimenez, Melanie and Corrigan, Damion K and Patel, Bhavik Anil},
title = {Development of a High-Sensitivity Electrochemical Immunoassay Using a Fully 3D-Printed Electrocatalytic Microelectrode Probe Platform},
journal = {Analytical Chemistry},
year = {2026},
volume = {98},
number = {12},
pages = {9260--9272},
doi = {10.1021/acs.analchem.5c08037}
}
EUVIMED