EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices

Jessica L Seidelman iD, Arthur W Baker iD, Sarah S Lewis, Bobby G Warren iD, Aaron Barrett, Amanda Graves, Carly King, Bonnie Taylor iD, Jill Engel iD, Desiree Bonnadonna, Carmelo Milano, Richard J Wallace und 3 weitere

Infection Control & Hospital Epidemiology · 2024

DOI: 10.1017/ice.2023.273 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Infection Control & Hospital Epidemiology, 2024. Verfasst von Jessica L Seidelman und 14 weiteren Autor:innen. Beteiligte Einrichtung: Duke Center for Antimicrobial Stewardship and Infection Prevention, Durham, North Carolina. Hauptthemen laut MEDLINE: Mycobacterium Infections, Nontuberculous, Cross Infection, Mycobacterium Infections. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIAID NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Various water-based heater-cooler devices (HCDs) have been implicated in nontuberculous mycobacteria outbreaks. Ongoing rigorous surveillance for healthcare-associated M. abscessus (HA-Mab) put in place following a prior institutional outbreak of M. abscessus alerted investigators to a cluster of 3 extrapulmonary M. abscessus infections among patients who had undergone cardiothoracic surgery.

Methods

Investigators convened a multidisciplinary team and launched a comprehensive investigation to identify potential sources of M. abscessus in the healthcare setting. Adherence to tap water avoidance protocols during patient care and HCD cleaning, disinfection, and maintenance practices were reviewed. Relevant environmental samples were obtained. Patient and environmental M. abscessus isolates were compared using multilocus-sequence typing and pulsed-field gel electrophoresis. Smoke testing was performed to evaluate the potential for aerosol generation and dispersion during HCD use. The entire HCD fleet was replaced to mitigate continued transmission.

Results

Clinical presentations of case patients and epidemiologic data supported intraoperative acquisition. M. abscessus was isolated from HCDs used on patients and molecular comparison with patient isolates demonstrated clonality. Smoke testing simulated aerosolization of M. abscessus from HCDs during device operation. Because the HCD fleet was replaced, no additional extrapulmonary HA-Mab infections due to the unique clone identified in this cluster have been detected.

Conclusions

Despite adhering to HCD cleaning and disinfection strategies beyond manufacturer instructions for use, HCDs became colonized with and ultimately transmitted M. abscessus to 3 patients. Design modifications to better contain aerosols or filter exhaust during device operation are needed to prevent NTM transmission events from water-based HCDs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Mycobacterium Infections, NontuberculousCross InfectionMycobacterium InfectionsHumansNontuberculous MycobacteriaMultilocus Sequence TypingDisease Outbreaks

Autor:innen und Institutionen

  • Duke Center for Antimicrobial Stewardship and Infection Prevention, Durham, North Carolina.
  • Division of Infectious Diseases, Duke University Medical Center, Durham, North Carolina.
  • Disinfection, Resistance, Transmission and Epidemiology Laboratory, Department of Medicine, Duke University Medical Center, Durham, North Carolina.
  • Division of Cardiovascular and Thoracic Surgery, Duke University School of Medicine, Durham, North Carolina.
  • Mycobacteria/Nocardia Laboratory, University of Texas Health Science Center, Tyler, Texas.
  • Occupational and Environmental Safety Office, Laboratory Safety, Duke University and Health System, Durham, North Carolina.

ORCID-Profile: Jessica L Seidelman, Arthur W Baker, Bobby G Warren, Bonnie Taylor, Jill Engel, Deverick J Anderson

Förderung

  • NIAID NIH HHS

Zitieren

APA

Seidelman, J. L., Baker, A. W., Lewis, S. S., Warren, B. G., Barrett, A., Graves, A., King, C., Taylor, B., Engel, J., Bonnadonna, D., Milano, C., Wallace, R. J., Stiegel, M., Anderson, D. J., & Smith, B. A. (2024). A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices. Infection Control & Hospital Epidemiology. https://doi.org/10.1017/ice.2023.273

Vancouver

Seidelman JL, Baker AW, Lewis SS, Warren BG, Barrett A, Graves A, et al. A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices. Infection Control & Hospital Epidemiology. 2024. doi:10.1017/ice.2023.273

BibTeX
@article{Seidelman2024,
  author = {Seidelman, Jessica L and Baker, Arthur W and Lewis, Sarah S and Warren, Bobby G and Barrett, Aaron and Graves, Amanda and King, Carly and Taylor, Bonnie and Engel, Jill and Bonnadonna, Desiree and Milano, Carmelo and Wallace, Richard J and Stiegel, Matthew and Anderson, Deverick J and Smith, Becky A},
  title = {A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices},
  journal = {Infection Control & Hospital Epidemiology},
  year = {2024},
  doi = {10.1017/ice.2023.273}
}
Kostenlos starten →