VergleichsstudieOpen Access · CC BY 4.0

Comparison of PetMAP digital and Mindray wrist oscillometric blood pressure measurements with invasive blood pressure in anaesthetized bonobos (Pan paniscus)

Philippa Bucknell, Florence Hillen, Phillipa Dobbs, Samantha Ashfield, Kate White

Veterinary Anaesthesia and Analgesia · 2026

DOI: 10.1016/j.vaa.2025.10.009 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Veterinary Anaesthesia and Analgesia, 2026. Verfasst von Philippa Bucknell und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Veterinary Medicine and Science, University of Nottingham, Sutton Bonington Campus, Loughborough, UK; Twycross Zoo, Atherstone, UK. Hauptthemen laut MEDLINE: Blood Pressure Determination, Pan paniscus, Blood Pressure, Anesthesia. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To evaluate the agreement of the PetMAP noninvasive blood pressure (NIBP) device with digital cuff and the Mindray NIBP device with wrist cuff with invasive blood pressure (IBP) in bonobos. STUDY DESIGN: Prospective clinical study (non-blinded or -randomized). ANIMALS: A group of 12 bonobos aged 5-44 years undergoing anaesthesia for enclosure movement and health checks. METHODS: Systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressure measurements were recorded simultaneously at 5-minute intervals using the NIBP devices and IBP via anterior tibial artery cannulation. Modified Bland-Altman and adapted error grid analyses assessed agreement between IBP and NIBP and the clinical implications of any biases respectively. A 95% limit of agreement (LOA) estimated the possible magnitude of sampling error. Pearson r correlations between IBP and NIBP measurements were calculated. A p < 0.05 was considered significant. RESULTS: IBP ranges were SAP 50-137 mmHg, MAP 32-85 mmHg and DAP 20-69 mmHg. Biases with the PetMAP were -6.8 mmHg for SAP (LOA -41.3 to 27.7 mmHg), -7.2 mmHg for MAP (LOA -30.7 to 16.3 mmHg) and -3.1 mmHg for DAP (LOA -25.5 to 19.3 mmHg) from 84 PetMAP-IBP paired measurements. Correlation coefficients were 0.6 for SAP and DAP and 0.7 for MAP. Whilst 70% of MAP values were clinically acceptable, 13% demonstrated unrecognised hypotension. The Mindray demonstrated biases of -25.3 mmHg for SAP (LOA -69.5 to 18.9 mmHg), -17.5 mmHg for MAP (LOA -51.4 to 16.5 mmHg) and -14.4 mmHg for DAP (LOA -42.2 to 13.5 mmHg) from 95 Mindray-IBP paired measurements. Correlation coefficients were 0.7 for all. Only 45% of MAP values were clinically acceptable and 26% demonstrated unrecognised hypotension. CONCLUSIONS AND CLINICAL RELEVANCE: Whilst the PetMAP agreed more closely with IBP, neither device met all the American College of Veterinary Internal Medicine validation criteria, and neither should be used to identify hypotension.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Pressure DeterminationPan paniscusBlood PressureAnesthesiaBlood Pressure MonitorsAnimalsFemaleMaleProspective Studies

Stichwörter der Autor:innen

apebonobocardiovascularinvasive blood pressurenoninvasive blood pressurevalidation

Autor:innen und Institutionen

  • School of Veterinary Medicine and Science, University of Nottingham, Sutton Bonington Campus, Loughborough, UK; Twycross Zoo, Atherstone, UK.
  • School of Veterinary Medicine and Science, University of Nottingham, Sutton Bonington Campus, Loughborough, UK.
  • Twycross Zoo, Atherstone, UK.
  • School of Veterinary Medicine and Science, University of Nottingham, Sutton Bonington Campus, Loughborough, UK. Electronic address: Kate.white@nottingham.ac.uk.

Zitieren

APA

Bucknell, P., Hillen, F., Dobbs, P., Ashfield, S., & White, K. (2026). Comparison of PetMAP digital and Mindray wrist oscillometric blood pressure measurements with invasive blood pressure in anaesthetized bonobos (Pan paniscus). Veterinary Anaesthesia and Analgesia. https://doi.org/10.1016/j.vaa.2025.10.009

Vancouver

Bucknell P, Hillen F, Dobbs P, Ashfield S, White K. Comparison of PetMAP digital and Mindray wrist oscillometric blood pressure measurements with invasive blood pressure in anaesthetized bonobos (Pan paniscus). Veterinary Anaesthesia and Analgesia. 2026. doi:10.1016/j.vaa.2025.10.009

BibTeX
@article{Bucknell2026,
  author = {Bucknell, Philippa and Hillen, Florence and Dobbs, Phillipa and Ashfield, Samantha and White, Kate},
  title = {Comparison of PetMAP digital and Mindray wrist oscillometric blood pressure measurements with invasive blood pressure in anaesthetized bonobos (Pan paniscus)},
  journal = {Veterinary Anaesthesia and Analgesia},
  year = {2026},
  doi = {10.1016/j.vaa.2025.10.009}
}
Kostenlos starten →