Correlating Infrared Surface Thermometry With Core Temperature Assessments in Simulated Cooling and Rewarming of Porcine Kidneys
Transplantation Proceedings · 2026-07 · 58 (6) · 1201-1206
DOI: 10.1016/j.transproceed.2026.05.019 ↗Kurz gefasst
Fachartikel in der Zeitschrift Transplantation Proceedings, 2026-07. Verfasst von Martin Igbokwe und 7 weiteren Autor:innen. Beteiligte Einrichtung: Multi-Organ Transplant Program, London Health Sciences Centre, London, Ontario, Canada; Schulich School of Medicine and Dentistry, London, Ontario, C…. Hauptthemen laut MEDLINE: Kidney, Rewarming, Body Temperature, Infrared Rays. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 13 Literaturangaben. Gefördert von: Physicians' Services Inc Foundation.
Abstract (Originalsprache)
BACKGROUND In developing strategies to analyze the effects of controlled rewarming of donor kidneys, accurate and reproducible measurements of renal temperatures are required. While invasive probes provide accurate core temperature measurements, they risk tissue damage and may compromise the sterile field. Infrared thermometry (IRT) offers a non-contact alternative, but correlation with renal core temperature remains insufficiently validated in transplant-related scenarios. OBJECTIVE To correlate surface IRT temperatures compared with core temperature assessments in porcine models of donor kidney cooling and rewarming. METHODS This study was conducted in a controlled surgical setting. Porcine kidneys were procured for study. Core temperatures were measured with a calibrated thermocouple probe (Physitemp Thermalert TH-8) inserted into the renal pelvis via the ureter. Surface temperature was recorded using a non-contact infrared thermometer (CEM DT-8839) held 10 to 15 cm perpendicular to the renal capsule. Measurements were obtained by both techniques every minute from cold flush with perfusion solution and during rewarming. Agreement between modalities was assessed with Bland-Altman analysis, and correlation was determined using Pearson's coefficient. RESULTS Across all phases, IRT measurements demonstrated strong correlation with core temperatures (r > 0.9, p < .001). Bland-Altman plots indicated minimal systematic bias, with narrower limits of agreement during steady-state cooling and wider variability during rewarming. IRT effectively tracked temperature trends, though minor deviations (±0.5-1.0°C) from core readings were observed during dynamic thermal transitions. CONCLUSIONS Surface IRT has strong correlation with core temperature during pig kidney cooling and rewarming. IRT is well suited for trend monitoring and intraoperative workflow guidance during evaluation of controlled rewarming of donor kidneys.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Multi-Organ Transplant Program, London Health Sciences Centre, London, Ontario, Canada; Schulich School of Medicine and Dentistry, London, Ontario, Canada.
- Matthew Mailing Center for Translational Transplant Studies, London Health Sciences Centre, Western University, London, Ontario, Canada; Schulich School of Medicine and Dentistry, London, Ontario, Canada.
- Schulich School of Medicine and Dentistry, London, Ontario, Canada.
- Matthew Mailing Center for Translational Transplant Studies, London Health Sciences Centre, Western University, London, Ontario, Canada.
- Multi-Organ Transplant Program, London Health Sciences Centre, London, Ontario, Canada; Matthew Mailing Center for Translational Transplant Studies, London Health Sciences Centre, Western University, London, Ontario, Ca…
Förderung
- Physicians' Services Inc Foundation
Zitieren
APA
Igbokwe, M., Wang, J., Byrne, C., Vyas, K. A., Richard-Mohammed, M., Alfaifi, A., Sener, A., & Luke, P. P. (2026). Correlating Infrared Surface Thermometry With Core Temperature Assessments in Simulated Cooling and Rewarming of Porcine Kidneys. Transplantation Proceedings, 58(6), 1201-1206. https://doi.org/10.1016/j.transproceed.2026.05.019
Vancouver
Igbokwe M, Wang J, Byrne C, Vyas KA, Richard-Mohammed M, Alfaifi A, et al. Correlating Infrared Surface Thermometry With Core Temperature Assessments in Simulated Cooling and Rewarming of Porcine Kidneys. Transplantation Proceedings. 2026;58(6):1201-1206. doi:10.1016/j.transproceed.2026.05.019
BibTeX
@article{Igbokwe2026,
author = {Igbokwe, Martin and Wang, John and Byrne, Cory and Vyas, Khushi A and Richard-Mohammed, Mahmoud and Alfaifi, Abdullah and Sener, Alp and Luke, Patrick P},
title = {Correlating Infrared Surface Thermometry With Core Temperature Assessments in Simulated Cooling and Rewarming of Porcine Kidneys},
journal = {Transplantation Proceedings},
year = {2026},
volume = {58},
number = {6},
pages = {1201--1206},
doi = {10.1016/j.transproceed.2026.05.019}
}
EUVIMED