PGMI assessment in mammography: AI software versus human readers

T Santner, C Ruppert, S Gianolini, J-G Stalheim, S Frei, M Hondl, V Fröhlich, S Hofvind, G Widmann

Radiography · 2025

DOI: 10.1016/j.radi.2025.103017 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Radiography, 2025. Verfasst von T Santner und 8 weiteren Autor:innen. Beteiligte Einrichtung: Medical University of Innsbruck, Fritz-Pregl-Strasse 3, 6020, Innsbruck, Austria. Electronic address: tina.santner@student.i-med.ac.at. Hauptthemen laut MEDLINE: Mammography, Software, Artificial Intelligence, Breast Neoplasms. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: The aim of this study was to evaluate human inter-reader agreement of parameters included in PGMI (perfect-good-moderate-inadequate) classification of screening mammograms and explore the role of artificial intelligence (AI) as an alternative reader. METHODS: Five radiographers from three European countries independently performed a PGMI assessment of 520 anonymized mammography screening examinations randomly selected from representative subsets from 13 imaging centres within two European countries. As a sixth reader, a dedicated AI software was used. Accuracy, Cohen's Kappa, and confusion matrices were calculated to compare the predictions of the software against the individual assessment of the readers, as well as potential discrepancies between them. A questionnaire and a personality test were used to better understand the decision-making processes of the human readers. RESULTS: Significant inter-reader variability among human readers with poor to moderate agreement (κ = -0.018 to κ = 0.41) was observed, with some showing more homogenous interpretations of single features and overall quality than others. In comparison, the software surpassed human inter-reader agreement in detecting glandular tissue cuts, mammilla deviation, pectoral muscle detection, and pectoral angle measurement, while remaining features and overall image quality exhibited comparable performance to human assessment. CONCLUSION: Notably, human inter-reader disagreement of PGMI assessment in mammography is considerably high. AI software may already reliably categorize quality. Its potential for standardization and immediate feedback to achieve and monitor high levels of quality in screening programs needs further attention and should be included in future approaches. IMPLICATIONS FOR PRACTICE: AI has promising potential for automated assessment of diagnostic image quality. Faster, more representative and more objective feedback may support radiographers in their quality management processes. Direct transformation of common PGMI workflows into an AI algorithm could be challenging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MammographySoftwareArtificial IntelligenceBreast NeoplasmsRadiographic Image Interpretation, Computer-AssistedHumansFemaleObserver VariationEuropeBreastMiddle Aged

Stichwörter der Autor:innen

AIImage qualityMammographyPGMIQuality assessmentSoftware

Autor:innen und Institutionen

  • Medical University of Innsbruck, Fritz-Pregl-Strasse 3, 6020, Innsbruck, Austria. Electronic address: tina.santner@student.i-med.ac.at.
  • b-rayZ AG, Wagistrasse 21, 8952, Schlieren, Switzerland. Electronic address: carlotta.ruppert@b-rayz.ch.
  • MSS Medical Software Solutions GmbH, Holzwiesstrasse 48, 8703, Erlenbach, Switzerland. Electronic address: stefano@gianolini.ch.
  • Evidia, Kaigaten 5, 5015, Bergen, Norway. Electronic address: johanne.gro.stalheim@evidia.no.
  • Way to Women Sàrl, Chemin du Pré de l'Epine 7, 1261, Le Vaud, Switzerland. Electronic address: w4women@outlook.com.
  • Klinik Ottakring, Montleartstraße 37, 1160, Wien, Austria. Electronic address: michaela.hondl-adametz@gesundheitsverbund.at.

Zitieren

APA

Santner, T., Ruppert, C., Gianolini, S., Stalheim, J. G., Frei, S., Hondl, M., Fröhlich, V., Hofvind, S., & Widmann, G. (2025). PGMI assessment in mammography: AI software versus human readers. Radiography. https://doi.org/10.1016/j.radi.2025.103017

Vancouver

Santner T, Ruppert C, Gianolini S, Stalheim JG, Frei S, Hondl M, et al. PGMI assessment in mammography: AI software versus human readers. Radiography. 2025. doi:10.1016/j.radi.2025.103017

BibTeX
@article{Santner2025,
  author = {Santner, T and Ruppert, C and Gianolini, S and Stalheim, J-G and Frei, S and Hondl, M and Fröhlich, V and Hofvind, S and Widmann, G},
  title = {PGMI assessment in mammography: AI software versus human readers},
  journal = {Radiography},
  year = {2025},
  doi = {10.1016/j.radi.2025.103017}
}
Kostenlos starten →