Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose
Radiography · 2024
DOI: 10.1016/j.radi.2024.04.016 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiography, 2024. Verfasst von R Nelson und 2 weiteren Autor:innen. Beteiligte Einrichtung: Radiology Department, Jubilee Wing, Leeds General Infirmary, Leeds, LS1 3EX, UK. Electronic address: ruthnelson@nhs.net. Hauptthemen laut MEDLINE: Patient Positioning, Radiation Dosage, Tomography, X-Ray Computed, Imaging, Three-Dimensional. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
INTRODUCTION: 3D positioning cameras that automate the positioning of patients with respect to the CT isocentre have been developed and are in common use in CT departments. This study aimed to compare the performance of radiographers and a 3D camera system with respect to positioning accuracy and the effect on patient radiation dose for chest-abdomen-pelvis scans. METHODS: Patient positioning and dose data obtained from a dose management system was evaluated over a two-month period for patients positioned with (CAMon) and without (CAMoff) the positioning camera. Median vertical and lateral offset values were compared between the groups whilst doses were evaluated as a function of patient water equivalent diameter (WED) for the thorax and abdomen-pelvis acquisitions for both cohorts. RESULTS: Radiographers demonstrated high levels of positioning accuracy, however significant improvements in median vertical offset were identified for the CAMon cohort for both thorax (8 mm vs. 17 mm (p = 0.001)) and abdomen-pelvis (7 mm vs. 16 mm (p = 0.003)) scans. The percentage of patients positioned within 5 mm of the isocentre was 39.0% and 16.1% for the CAMon and CAMoff cohorts. For CAMoff scans, 77.4% of patients were positioned below the isocentre, but this was reduced to 45.8% for CAMon scans. No significant changes in dose as a function of WED were identified related to the camera use (thorax: p = 0.569, abdomen-pelvis: p = 0.760). CONCLUSION: Use of a 3D camera delivered significant improvements in the accuracy and reproducibility of patient positioning when compared with radiographers. IMPLICATIONS FOR PRACTICE: Improvements in positioning accuracy were observed at the research site and hence positioning camera use has the potential to become standard practice in CT to help ensure appropriate doses are delivered to patients according to their size.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiology Department, Jubilee Wing, Leeds General Infirmary, Leeds, LS1 3EX, UK. Electronic address: ruthnelson@nhs.net.
- Radiology Department, Jubilee Wing, Leeds General Infirmary, Leeds, LS1 3EX, UK. Electronic address: scott.harkin@nhs.net.
- Faculty of Health Studies, University of Bradford, Bradford, BD7 1DP, UK. Electronic address: g.r.iball@bradford.ac.uk.
Zitieren
APA
Nelson, R., Harkin, S., & Iball, G. (2024). Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose. Radiography. https://doi.org/10.1016/j.radi.2024.04.016
Vancouver
Nelson R, Harkin S, Iball G. Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose. Radiography. 2024. doi:10.1016/j.radi.2024.04.016
BibTeX
@article{Nelson2024,
author = {Nelson, R and Harkin, S and Iball, G},
title = {Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose},
journal = {Radiography},
year = {2024},
doi = {10.1016/j.radi.2024.04.016}
}
EUVIMED