Fallbericht in der Zeitschrift The Journal of Prosthetic Dentistry, 2026. Verfasst von Yi He und 5 weiteren Autor:innen. Beteiligte Einrichtung: Master's student, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, …. Hauptthemen laut MEDLINE: Titanium, Peri-Implantitis, Dental Implants. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Kunming Medical University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The release of titanium particles is a common yet often overlooked complication of implant therapy. Most research has focused on nanoscale or submicrometer titanium particles and their immunological effects, with limited data on the long-term tissue response to larger particles. This report presents a patient with a mechanical implant complication accompanied by peri-implantitis. The improper removal of a fractured abutment screw resulted in severe implant damage and the release of numerous 20- to 200-µm titanium particles, which remained in the surrounding tissue for 9 years. After implant removal, scanning electron microscopy, energy-dispersive X-ray spectroscopy, histopathology, and immunohistochemistry were used to evaluate the tissue response. The findings showed that long-standing micrometer-sized titanium particles were primarily encapsulated by fibrous tissue rather than eliciting an inflammatory reaction. This report provides clinical evidence that the prolonged presence of micrometer-sized titanium particles in tissues does not necessarily trigger a significant immune or inflammatory response.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TitaniumPeri-ImplantitisDental ImplantsHumansCorrosionIatrogenic DiseaseMicroscopy, Electron, ScanningMiddle AgedMaleDental Restoration FailureFemaleDevice RemovalParticle Size

Substanzen

TitaniumDental Implants

Autor:innen und Institutionen

  • Master's student, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, Kunming, PR China; and Master's student, Yunnan Key Laboratory of Stom…
  • PhD candidate, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, Kunming, PR China; and PhD candidate, Yunnan Key Laboratory of Stomatolog…
  • Associate Professor, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, Kunming, PR China.
  • Master's student, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, Kunming, PR China.
  • Professor, Department of Oral Surgery, 920th Hospital of Joint Logistics Support Force, PLA, Teaching Hospital of Kunming Medical University, Kunming, PR China; and Professor, Yunnan Key Laboratory of Stomatology and De…

Förderung

  • Kunming Medical University

Zitieren

APA

He, Y., Wang, D. J., Hu, Y. T., Zhao, X. L., Jiang, Y., & Xiao, Y. H. (2026). Peri-implantitis with iatrogenic implant corrosion: A clinical report of titanium particles remaining in local tissues for up to nine years. The Journal of Prosthetic Dentistry. https://doi.org/10.1016/j.prosdent.2025.04.024

Vancouver

He Y, Wang DJ, Hu YT, Zhao XL, Jiang Y, Xiao YH. Peri-implantitis with iatrogenic implant corrosion: A clinical report of titanium particles remaining in local tissues for up to nine years. The Journal of Prosthetic Dentistry. 2026. doi:10.1016/j.prosdent.2025.04.024

BibTeX
@article{He2026,
  author = {He, Yi and Wang, Ding-Jie and Hu, Yun-Tong and Zhao, Xi-Long and Jiang, Ya and Xiao, Yu-Hong},
  title = {Peri-implantitis with iatrogenic implant corrosion: A clinical report of titanium particles remaining in local tissues for up to nine years},
  journal = {The Journal of Prosthetic Dentistry},
  year = {2026},
  doi = {10.1016/j.prosdent.2025.04.024}
}
Kostenlos starten →