Influence of different designs of bulb support structures on the accuracy and weight of 3D printed maxillary obturators. An in vitro study
The Journal of Prosthetic Dentistry · 2024
DOI: 10.1016/j.prosdent.2024.04.023 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Prosthetic Dentistry, 2024. Verfasst von Reham B Osman und 1 weiteren Autor:innen. Beteiligte Einrichtung: Associate Professor, Prosthodontics Department, Faculty of Dentistry, Cairo University, Cairo, Egypt. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Palatal Obturators. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
STATEMENT OF PROBLEM: Evidence is lacking on the influence of different designs of bulb support structures on the accuracy of 3-dimensional (3D) printed obturators. PURPOSE: The purpose of this in vitro study was to evaluate the influence of various infill designs (hollow, honeycomb, and gyroid) for the bulb of an obturator on the accuracy and weight of digital light projection (DLP) 3D printed maxillary obturators. MATERIAL AND METHODS: A maxillary obturator was virtually designed and used to obtain 3 digital reference files which were defined based on the design of the infill support structure within the bulb: hollow, honeycomb, and gyroid. The resultant standard tessellation language (STL) files were used to fabricate the obturators using a DLP 3D-printer in FREEPRINT denture resin material (n=10 per each group design). The fitting surfaces of all printed specimens (n=30) were digitized using a laboratory scanner, and the scan STL files were exported to the Geomagic control X program for dimensional accuracy analysis (trueness and precision) using the digital subtraction technique. One-way analysis of variance (ANOVA) was used for analysis (α=.05) RESULTS: Quantitative analysis revealed no significant difference in root mean square estimate (RMSE) values among the test groups for trueness (P=.326) and precision (P=.140). Hollow bulb design was significantly lighter in weight than both honeycomb and gyroid infill designs (P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Associate Professor, Prosthodontics Department, Faculty of Dentistry, Cairo University, Cairo, Egypt.
- Associate Professor, Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia. Electronic address: nalharbi@ksu.edu.sa.
Zitieren
APA
Osman, R. B., & Alharbi, N. (2024). Influence of different designs of bulb support structures on the accuracy and weight of 3D printed maxillary obturators. An in vitro study. The Journal of Prosthetic Dentistry. https://doi.org/10.1016/j.prosdent.2024.04.023
Vancouver
Osman RB, Alharbi N. Influence of different designs of bulb support structures on the accuracy and weight of 3D printed maxillary obturators. An in vitro study. The Journal of Prosthetic Dentistry. 2024. doi:10.1016/j.prosdent.2024.04.023
BibTeX
@article{Osman2024,
author = {Osman, Reham B and Alharbi, Nawal},
title = {Influence of different designs of bulb support structures on the accuracy and weight of 3D printed maxillary obturators. An in vitro study},
journal = {The Journal of Prosthetic Dentistry},
year = {2024},
doi = {10.1016/j.prosdent.2024.04.023}
}