Accuracy of 3 calibration methods of computer-assisted dynamic navigation for implant placement: An in vitro study

Xiyan Pei, Xiaoqiang Liu, Siniong Iao, Feifei Ma, Hong Li, Feng Sun

The Journal of Prosthetic Dentistry · 2024-04 · 131 (4) · 668-674

DOI: 10.1016/j.prosdent.2022.03.014 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Prosthetic Dentistry, 2024-04. Verfasst von Xiyan Pei und 5 weiteren Autor:innen. Beteiligte Einrichtung: Lecturer, First Clinic Division, Peking University School and Hospital of Stomatology & National Center of Stomatology & National Clinical Research C…. Hauptthemen laut MEDLINE: Dental Implants, Surgery, Computer-Assisted, Mouth, Edentulous. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 34 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

STATEMENT OF PROBLEM: Dynamic navigation for implant placement has been reported to be more accurate than freehand surgery. However, the accuracy of the calibration methods used for navigation in partially edentulous individuals with distal extensions remains unknown. PURPOSE: The purpose of this in vitro study on dental models was to evaluate the accuracy of 3 calibration methods of dynamic navigation for implant placement in the distal extension of partially edentulous arches. MATERIAL AND METHODS: Eleven standardized polyurethane mandibular models with distal extensions were prepared. The left first molar, second molar, and second premolar from each model (33 tooth sites) were randomly assigned to 1 of the 3 calibration methods: U-shaped tube embedded with radiopaque markers, anatomic tooth cusps, and bone markers with the random number table method. Preoperative and postoperative cone beam computed tomography images were obtained for deviation analyses. The primary outcomes were 3-dimensional (3D) deviation at the implant platform and apex and angular deviation. Differences among the test groups were analyzed by using a 1-way analysis of variance (ANOVA) and the least significant difference (LSD) post hoc test (α=.05). RESULTS: The mean ±standard deviation 3D deviations were 0.78 ±0.34, 1.86 ±0.91, and 1.44 ±0.57 mm at the implant platform and 0.79 ±0.35, 2.19 ±1.01, and 1.49 ±0.50 mm at the apex in the U-shaped tube, tooth cusp, and bone marker groups, respectively. The 3D deviations at the implant platform and apex were significantly different among the groups (P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental ImplantsSurgery, Computer-AssistedMouth, EdentulousHumansDental Implantation, EndosseousCalibrationCone-Beam Computed TomographyComputersComputer-Aided DesignImaging, Three-Dimensional

Substanzen

Dental Implants

Autor:innen und Institutionen

  • Lecturer, First Clinic Division, Peking University School and Hospital of Stomatology & National Center of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Research Center of Oral…
  • Associate Professor, Department of Prosthodontics, Peking University School and Hospital of Stomatology, Beijing, PR China.
  • PhD student, Department of Periodontology, Peking University School and Hospital of Stomatology, Beijing, PR China.
  • Lecturer, First Clinic Division, Peking University School and Hospital of Stomatology, Beijing, PR China.
  • Clinical Professor, First Clinic Division, Peking University School and Hospital of Stomatology & National Center of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Research Cent…

Zitieren

APA

Pei, X., Liu, X., Iao, S., Ma, F., Li, H., & Sun, F. (2024). Accuracy of 3 calibration methods of computer-assisted dynamic navigation for implant placement: An in vitro study. The Journal of Prosthetic Dentistry, 131(4), 668-674. https://doi.org/10.1016/j.prosdent.2022.03.014

Vancouver

Pei X, Liu X, Iao S, Ma F, Li H, Sun F. Accuracy of 3 calibration methods of computer-assisted dynamic navigation for implant placement: An in vitro study. The Journal of Prosthetic Dentistry. 2024;131(4):668-674. doi:10.1016/j.prosdent.2022.03.014

BibTeX
@article{Pei2024,
  author = {Pei, Xiyan and Liu, Xiaoqiang and Iao, Siniong and Ma, Feifei and Li, Hong and Sun, Feng},
  title = {Accuracy of 3 calibration methods of computer-assisted dynamic navigation for implant placement: An in vitro study},
  journal = {The Journal of Prosthetic Dentistry},
  year = {2024},
  volume = {131},
  number = {4},
  pages = {668--674},
  doi = {10.1016/j.prosdent.2022.03.014}
}
Kostenlos starten →