Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction

M. Ahmad, J. Dickmann, N.A. Lackner, O.J. Ott, R. Fietkau, S. Corradini, C. Bert, J. Szkitsak

Physics and Imaging in Radiation Oncology · 2026-11 · 42 · 101126

DOI: 10.1016/j.phro.2026.101126 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics and Imaging in Radiation Oncology, 2026-11. Verfasst von M. Ahmad und 7 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose: Four-dimensional computed tomography (4DCT) is essential for motion-resolved imaging in radiotherapy, but stack alignment artifacts caused by irregular breathing may compromise anatomical accuracy. This study established a quantitative framework to evaluate reconstruction differences between clinical reference reconstruction (CRR) and stack artifact-reduced reconstruction (SARR) and identify clinically relevant changes. Materials and methods: Retrospective 4DCT datasets from 80 lung cancer patients were analyzed, comprising a rated cohort ( n = 30) and an independent control cohort ( n = 50). Geometric, intensity-based, and combined metrics, including gamma analysis, were calculated across ten respiratory phases for segmented organs-at-risk. Correlations with reader ratings identified metrics that best reflected clinically relevant reconstruction differences.

Results

Most similarity and intensity-based measures showed limited association with image quality differences (Pearson correlation coefficient r 0.3 reaching r ≈ 0.59 for phase-based data (95% confidence interval 0.30–0.79). Reconstruction differences were predominantly observed in the liver and lower lung lobes, consistent with respiratory motion, while the spinal cord showed minimal variation. Robust z-score thresholding enabled automated categorization of moderate and substantial deviations. In the control cohort, 12% showed substantial and 14% moderate changes for phase-based reconstructions. In the patient with the largest changes, the liver appeared as a double contour in the CRR and as a single contour in the SARR.

Conclusions

Gamma-based evaluation provided an objective method to detect and localize reconstruction differences in 4DCT, supporting transparent clinical implementation of advanced reconstruction algorithms in radiotherapy.

Zitieren

APA

Ahmad, M., Dickmann, J., Lackner, N., Ott, O., Fietkau, R., Corradini, S., Bert, C., & Szkitsak, J. (2026). Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction. Physics and Imaging in Radiation Oncology, 42, 101126. https://doi.org/10.1016/j.phro.2026.101126

Vancouver

Ahmad M, Dickmann J, Lackner N, Ott O, Fietkau R, Corradini S, et al. Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction. Physics and Imaging in Radiation Oncology. 2026;42:101126. doi:10.1016/j.phro.2026.101126

BibTeX
@article{Ahmad2026,
  author = {Ahmad, M. and Dickmann, J. and Lackner, N.A. and Ott, O.J. and Fietkau, R. and Corradini, S. and Bert, C. and Szkitsak, J.},
  title = {Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction},
  journal = {Physics and Imaging in Radiation Oncology},
  year = {2026},
  volume = {42},
  pages = {101126},
  doi = {10.1016/j.phro.2026.101126}
}
Kostenlos starten →