A novel method of verifying appropriate nasogastric tube placement: animal and human studies

David Qing Wei Tan, Lien Khai Pang, Jasmine Sin Ee Kang, Caren Kar Yen Mok, Fiona Ai Ping Tan, Eunice Chew, Sanchalika Acharyya, Wern Hsien Bin, Ravinder Singh Sachdev, Nicole Kessa Xin-Hui Wee, Eric Chee Mun Loh, Khai Pang Leong

Nutrition · 2026

DOI: 10.1016/j.nut.2025.112933 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Nutrition, 2026. Verfasst von David Qing Wei Tan und 11 weiteren Autor:innen. Beteiligte Einrichtung: Nasotrak Medical Pte Ltd, Singapore. Hauptthemen laut MEDLINE: Intubation, Gastrointestinal, Enteral Nutrition. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

AIM: It is important to verify the correct placement of nasogastric tube (NGT) but this can be unreliable, difficult to perform, or prone to confounding factors. We developed a novel method that uses ultrasound to verify the location of a specially designed NGT in real time. We describe our findings using this system in both animal and first-in-human studies. METHODS: Our system consists of a portable hand-held device coupled with a disposable nasogastric tube embedded with a sensor at its tip. During operation, the device is connected electrically to the external portion of the tube and then held over the abdominal wall whereby it will emit audible and visible signal when in-situ tube tip is proximate. For proof of concept, we used an anaesthetized Yorkshire-Landrace pig. Three blinded observers scanned for the location of the nasogastric tube thrice, with the location altered between each. In the first-in-man study, we placed the tube in 12 healthy volunteers, and five blinded and independent observers scanned for the location of the tip. An X-ray was taken of each of the volunteers. RESULTS: In the animal study, the observers correctly identified the location of the nasogastric tube in every attempt. The mean time taken was 51.5 ± 7.53 seconds. In the human study, the nasogastric tube was detected 91.4% of the time when determined to be correctly placed in the stomach by the observers. The mean distance from the detected to actual NG tube tip was 6.5 cm (SD 6.9 cm). No complications occurred, except transient vomiting in one volunteer. CONCLUSION: This novel nasogastric tube verification system is accurate and potentially offers advantages over existing methods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intubation, GastrointestinalEnteral NutritionAnimalsHumansSwineMaleAdultFemaleUltrasonography

Stichwörter der Autor:innen

consistent with CINAHL/MeSH headings: nasogastric tubefeasibility studiesmedical device

Autor:innen und Institutionen

  • Nasotrak Medical Pte Ltd, Singapore.
  • Department of Nursing, Tan Tock Seng Hospital, Singapore.
  • Trendlines Medical Singapore Pte Ltd, Singapore.
  • Clinical Research and Innovation Office, Tan Tock Seng Hospital, Singapore.
  • Department of Anaesthesia, Khoo Teck Puat Hospital, Singapore.
  • Department of Continuing and Community Care, Tan Tock Seng Hospital, Singapore.

Zitieren

APA

Tan, D. Q. W., Pang, L. K., Kang, J. S. E., Mok, C. K. Y., Tan, F. A. P., Chew, E., Acharyya, S., Bin, W. H., Sachdev, R. S., Wee, N. K. X., Loh, E. C. M., & Leong, K. P. (2026). A novel method of verifying appropriate nasogastric tube placement: animal and human studies. Nutrition. https://doi.org/10.1016/j.nut.2025.112933

Vancouver

Tan DQW, Pang LK, Kang JSE, Mok CKY, Tan FAP, Chew E, et al. A novel method of verifying appropriate nasogastric tube placement: animal and human studies. Nutrition. 2026. doi:10.1016/j.nut.2025.112933

BibTeX
@article{Tan2026,
  author = {Tan, David Qing Wei and Pang, Lien Khai and Kang, Jasmine Sin Ee and Mok, Caren Kar Yen and Tan, Fiona Ai Ping and Chew, Eunice and Acharyya, Sanchalika and Bin, Wern Hsien and Sachdev, Ravinder Singh and Wee, Nicole Kessa Xin-Hui and Loh, Eric Chee Mun and Leong, Khai Pang},
  title = {A novel method of verifying appropriate nasogastric tube placement: animal and human studies},
  journal = {Nutrition},
  year = {2026},
  doi = {10.1016/j.nut.2025.112933}
}
Kostenlos starten →