Repeatability of in vivo MR cytometry for measuring cell size and density in healthy human livers

Xiaoyu Jiang, Chaoyang Jin, Hakmook Kang, Gracie Piantek, Zhongliang Zu, John C Gore, Junzhong Xu

Magnetic Resonance Imaging · 2026

DOI: 10.1016/j.mri.2025.110563 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2026. Verfasst von Xiaoyu Jiang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN 37232, USA; Department of Radiology and Radio…. Hauptthemen laut MEDLINE: Liver, Magnetic Resonance Imaging. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NCI NIH HHS, NIDDK NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To evaluate the repeatability of MRI-derived cell size and density measurements in the healthy human liver using MR cytometry, a non-invasive technique for quantifying tissue microstructure, METHODS: Eight healthy subjects underwent two identical MR cytometry scans (2-7 days apart) on a Phillips 3 T scanner. Diffusion data were acquired using pulsed gradient spin echo (PGSE) and oscillating gradient spin echo (OGSE) sequences at multiple diffusion times. Bayes factor model selection compared different MR cytometry signal models, such as IMPULSED, which ignores water exchange between intra and extracellular compartments, and JOINT, which accounts for it. Repeatability of MRI-derived parameters were assessed via whole-volume and multi-size ROI analyses using Bland-Altman plots and correlation coefficients. RESULTS: The JOINT model was decisively preferred (>90 % of voxels). High repeatability was observed for d, vin, ADCex, and cell density in whole-liver and larger ROIs, with correlations up to r = 0.96. ROI size significantly improved consistency. Coefficients of variation for cell density decreased from 72 % (single voxel) to 28 % (7 × 7 voxels). The exchange rate kin could not be reliably estimated. CONCLUSION: ROI-based MR cytometry provides repeatable measurements of mean cell size and density, supporting its potential for clinical applications such as assessing steatohepatitis and detecting hepatocellular carcinoma. Protocol optimization to improve SNR is needed for reliable quantification of water exchange.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LiverMagnetic Resonance ImagingHumansReproducibility of ResultsMaleAdultCell SizeFemaleCell CountMiddle AgedHealthy Volunteers

Stichwörter der Autor:innen

Cell densityCell sizeLiver microstructureMR cytometryOGSERepeatability

Autor:innen und Institutionen

  • Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN 37232, USA; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN…
  • Department of Biostatistics, Vanderbilt University Medical Center, Nashville, TN 37232, USA.
  • Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN 37232, USA.

Förderung

  • NCI NIH HHS
  • NIDDK NIH HHS
  • National Institutes of Health
  • National Institutes of Health National Cancer Institute
  • National Institute of Diabetes and Digestive and Kidney Diseases

Zitieren

APA

Jiang, X., Jin, C., Kang, H., Piantek, G., Zu, Z., Gore, J. C., & Xu, J. (2026). Repeatability of in vivo MR cytometry for measuring cell size and density in healthy human livers. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110563

Vancouver

Jiang X, Jin C, Kang H, Piantek G, Zu Z, Gore JC, et al. Repeatability of in vivo MR cytometry for measuring cell size and density in healthy human livers. Magnetic Resonance Imaging. 2026. doi:10.1016/j.mri.2025.110563

BibTeX
@article{Jiang2026,
  author = {Jiang, Xiaoyu and Jin, Chaoyang and Kang, Hakmook and Piantek, Gracie and Zu, Zhongliang and Gore, John C and Xu, Junzhong},
  title = {Repeatability of in vivo MR cytometry for measuring cell size and density in healthy human livers},
  journal = {Magnetic Resonance Imaging},
  year = {2026},
  doi = {10.1016/j.mri.2025.110563}
}
Kostenlos starten →