A human-compatible gradient coil for visualizing ultrasound propagation
Magnetic Resonance Imaging · 2025
DOI: 10.1016/j.mri.2025.110447 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2025. Verfasst von Paul-Emile Passe-Carlus und 5 weiteren Autor:innen. Beteiligte Einrichtung: Computer And Electrical Engineering, Brigham Young University, Provo, UT 84663, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NIBIB NIH HHS, NSF, National Science Foundation Division of Chemical Bioengineering Environmental and Transport Systems.
Abstract (Originalsprache)
In pursuit of a magnetic resonance imaging (MRI) based means to tailor transcranial focused ultrasound neuromodulation to a patient's unique skull morphology, this study presents a specialized gradient coil that sensitizes MRI images to ultrasonic vibrations at depths equivalent to the human cortex in a technique called the magnetic resonance hydrophone. The coil comprises a 60 mm diameter, pancake-style design that encodes acoustic displacements into MR images at the cost of an inhomogeneous encoding field. The coil was coupled with a 500 kHz, custom built, ultrasonic transducer. Both the magnetic field gradient of the coil and the acoustic field of the transducer were characterized in benchtop experiments. Acoustic standing waves were estimated in silico. Resulting MR images displayed a sinusoidal phase pattern modulated by both the transducer's acoustic field and the coil's magnetic field gradient. Acoustic pressures were estimated from the resulting images and compared to hydrophone measurements. The pancake-style coil produced a pressure measurement uncertainty pattern due to electronic noise that increased exponentially with depth. Uncertainty at locations between 0 and 30 mm of depth within a region approximately 10 mm wide scaled between approximately 20 kPa and 100 kPa. On average, the MRH underestimated the hydrophone by 12 kPa with the difference between the two following a standard deviation of 21 kPa.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Computer And Electrical Engineering, Brigham Young University, Provo, UT 84663, USA.
- Computer And Electrical Engineering, Brigham Young University, Provo, UT 84663, USA; Department of Radiology, University of Utah, Salt Lake City, UT 84112, USA.
- Department of Radiology, University of Utah, Salt Lake City, UT 84112, USA.
- Computer And Electrical Engineering, Brigham Young University, Provo, UT 84663, USA. Electronic address: spallen@byu.edu.
Förderung
- NIBIB NIH HHS
- NSF
- National Science Foundation Division of Chemical Bioengineering Environmental and Transport Systems
- Focused Ultrasound Foundation
- National Science Foundation
- National Institutes of Health
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Passe-Carlus, P. E., Cavinatto, D., Thyberg, N., Reed, C., Webb, T. D., & Allen, S. P. (2025). A human-compatible gradient coil for visualizing ultrasound propagation. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110447
Vancouver
Passe-Carlus PE, Cavinatto D, Thyberg N, Reed C, Webb TD, Allen SP. A human-compatible gradient coil for visualizing ultrasound propagation. Magnetic Resonance Imaging. 2025. doi:10.1016/j.mri.2025.110447
BibTeX
@article{PasseCarlus2025,
author = {Passe-Carlus, Paul-Emile and Cavinatto, Davi and Thyberg, Nathan and Reed, Carson and Webb, Taylor D and Allen, Steven P},
title = {A human-compatible gradient coil for visualizing ultrasound propagation},
journal = {Magnetic Resonance Imaging},
year = {2025},
doi = {10.1016/j.mri.2025.110447}
}
EUVIMED