Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique
Magnetic Resonance Imaging · 2025
DOI: 10.1016/j.mri.2025.110321 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2025. Verfasst von Thibault Willaume und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University Hospital of Strasbourg, 1 Avenue Moliere, 67098 Strasbourg, France. Electronic address: thibault.willaume@chru-st…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Imaging, Three-Dimensional, Cartilage, Articular, Knee Joint. Die Arbeit nennt 30 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: Compressed Sensing (CS) is an emerging technique to accelerate MRI acquisitions. The aim of this study was to assess the reliability and accuracy of cartilage thickness measurements in the knee using a CS-enabled isotropic 3D Fast Spin-Echo (FSE) sequence on a 3-T MRI scanner. METHODS: Twenty-eight tibial condyle sections were collected from 14 adult patients who underwent total knee arthroplasty. An isotropic 3D PDw FSE with variable flip-angle (CUBE) sequence with CS was used to acquire MR images of the tibial condyle sections. Minimum cartilage thickness measurements were independently performed by two experienced readers (R1 and R2) on MR images and compared to corresponding anatomical sections measurements. Intraclass correlation coefficients (ICCs) and Bland-Altman analyses, were used to assess agreement between MR and anatomical measurements. RESULTS: A total of 84 paired cartilage areas were analyzed [cartilage thickness measurements ranged from 0 to 3.40 mm at anatomical evaluation (mean, 1.08 mm ± 0.83)]. The agreements between MR and anatomical measurements were excellent (mean differences, 0.06 ± 0.31 mm for R1 and 0.03 ± 0.43 mm for R2) with respective ICC values of 0.93 and 0.88. Bland-Altman analyses revealed small differences between MR and anatomical measurements, with 95 % Limit of Agreements values falling within clinically acceptable ranges (-0.54 to 0.66 mm for R1, -0.87 to 0.80 mm for R2). CONCLUSION: The 3D PDw FSE sequence with Compressed Sensing acceleration technique demonstrated reliable and accurate assessment of cartilage thickness of the tibial condyles within a timeframe suitable for routine clinical practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University Hospital of Strasbourg, 1 Avenue Moliere, 67098 Strasbourg, France. Electronic address: thibault.willaume@chru-strasbourg.fr.
- Service de Chirurgie Orthopédique et de Traumatologie, Hôpital de Hautepierre, Hôpitaux Universitaires de Strasbourg, 1 avenue Molière, 67098 Strasbourg Cedex, France.
- Department of Pathology, University Hospital of Strasbourg, 67098 Strasbourg Cedex, France.
- Department of Radiology, University Hospital of Strasbourg, 1 Avenue Moliere, 67098 Strasbourg, France.
- Department of Radiology, University Hospital of Strasbourg, 1 Avenue Moliere, 67098 Strasbourg, France; ICube - UMR 7357, CNRS, University of Strasbourg, 67000 Strasbourg, France.
- Department of Interventional Radiology, Nouvel Hôpital Civil, 1 Place de L'hôpital, 67000 Strasbourg, France.
Zitieren
APA
Willaume, T., Ehlinger, M., Favreau, H., Weingertner, N., Zorn, P. E., Dillenseger, J. P., Koch, G., Velten, M., & Bierry, G. (2025). Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110321
Vancouver
Willaume T, Ehlinger M, Favreau H, Weingertner N, Zorn PE, Dillenseger JP, et al. Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique. Magnetic Resonance Imaging. 2025. doi:10.1016/j.mri.2025.110321
BibTeX
@article{Willaume2025,
author = {Willaume, Thibault and Ehlinger, Matthieu and Favreau, Henri and Weingertner, Noëlle and Zorn, Pierre-Emmanuel and Dillenseger, Jean-Philippe and Koch, Guillaume and Velten, Michel and Bierry, Guillaume},
title = {Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique},
journal = {Magnetic Resonance Imaging},
year = {2025},
doi = {10.1016/j.mri.2025.110321}
}
EUVIMED