Nitrided titanium alloy component enhances detection of small bone defects after total knee arthroplasty

Sho Masuda, Yukihide Minoda, Yohei Ohyama, Ryo Sugama, Yoichi Ohta, Hidetomi Terai

The Knee · 2025

DOI: 10.1016/j.knee.2025.07.014 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Knee, 2025. Verfasst von Sho Masuda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan. Hauptthemen laut MEDLINE: Titanium, Arthroplasty, Replacement, Knee, Knee Prosthesis, Femur. Die Arbeit nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: This study compared the accuracy of various imaging modalities in detecting small bone defects simulating radiolucent lines (RLLs) around femoral components made of nitrided titanium alloy (Ti-6Al-4V) versus cobalt-chromium (Co-Cr) alloy. METHODS: Eight femoral components (four nitrided titanium alloy and four Co-Cr alloy) were implanted into pig femurs. Two total knee arthroplasty (TKA) models were fabricated for each component: normal and RLL TKA models. Fluoroscopically guided plain radiography, tomosynthesis, computed tomography (CT), and magnetic resonance imaging (MRI) were performed. Thirty orthopaedic surgeons independently assessed the images to determine the presence or absence of bone defects. Sensitivity, specificity, and accuracy were compared between materials and imaging modalities. RESULTS: The accuracy of detecting bone defects was significantly higher in nitrided titanium alloy components than in Co-Cr alloy components using fluoroscopically guided plain radiography (66.7 % vs. 53.3 %, P = 0.048), tomosynthesis (87.5 % vs. 63.3 %, P < 0.001), and CT (100 % vs. 61.7 %, P < 0.001). MRI showed no significant difference (52.5 % vs. 55.8 %, P = 0.698). Furthermore, for nitrided titanium alloy components, the accuracy of detecting bone defects around the femoral components using CT was 100 %. CONCLUSIONS: Small bone defects simulating RLLs were more easily detected around nitrided titanium alloy femoral components than around Co-Cr alloy components. Nitrided titanium alloy femoral components may offer an advantage by enabling earlier detection of loosening and potentially reducing the need for complex revision surgery. These findings may serve as one of several factors for surgeons to consider when selecting implant materials for TKA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TitaniumArthroplasty, Replacement, KneeKnee ProsthesisFemurSwineAnimalsTomography, X-Ray ComputedChromium AlloysMagnetic Resonance ImagingFluoroscopyProsthesis DesignAlloys

Stichwörter der Autor:innen

Bone defectsEarly detectionNitrided titanium alloyRadiolucent lineTotal knee arthroplasty

Substanzen

TitaniumChromium Alloystitanium alloy (TiAl6V4)Alloys

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan.
  • Department of Orthopaedic Surgery, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan. Electronic address: yminoda@omu.ac.jp.
  • Department of Orthopaedic Surgery, Osaka Rosai Hospital, Osaka, Japan.

Zitieren

APA

Masuda, S., Minoda, Y., Ohyama, Y., Sugama, R., Ohta, Y., & Terai, H. (2025). Nitrided titanium alloy component enhances detection of small bone defects after total knee arthroplasty. The Knee. https://doi.org/10.1016/j.knee.2025.07.014

Vancouver

Masuda S, Minoda Y, Ohyama Y, Sugama R, Ohta Y, Terai H. Nitrided titanium alloy component enhances detection of small bone defects after total knee arthroplasty. The Knee. 2025. doi:10.1016/j.knee.2025.07.014

BibTeX
@article{Masuda2025,
  author = {Masuda, Sho and Minoda, Yukihide and Ohyama, Yohei and Sugama, Ryo and Ohta, Yoichi and Terai, Hidetomi},
  title = {Nitrided titanium alloy component enhances detection of small bone defects after total knee arthroplasty},
  journal = {The Knee},
  year = {2025},
  doi = {10.1016/j.knee.2025.07.014}
}
Kostenlos starten →