Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2025. Verfasst von Peter W Kurtz und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bioengineering, Clemson University, Clemson, South Carolina; The Clemson University-Medical University of South Carolina Bioengineering…. Hauptthemen laut MEDLINE: Knee Prosthesis, Arthroplasty, Replacement, Knee, Molybdenum, Cobalt. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
BACKGROUND: Total knee arthroplasty (TKA) device systems frequently include cast cobalt chrome alloy (CoCrMo) femoral components. However, compared to total hip arthroplasty (THA), gaps persist in our understanding of the correlations between femoral component damage, local metal release, and potential biological effects. Additionally, it remains unclear how TKA metal release affects clinical success or patient satisfaction. In this study, we investigated the associations between implant damage and metal release in the periprosthetic tissues following TKA. We asked: (1) does damage severity correlate with increased metal concentrations within the periprosthetic tissue? and (2) does the magnitude of metal released from CoCrMo femoral components merit clinical concern? METHODS: There were 51 CoCrMo femoral components and synovial samples that were prospectively collected by an institutional review board-exempt retrieval program. Devices received damage scores ranging from minimal (1) to severe (4). Tissue metal concentrations for cobalt (Co), chromium (Cr), molybdenum (Mo), and titanium (Ti) were quantified using inductively coupled plasma mass spectroscopy. RESULTS: Visual damage correlated with increases in Co, Cr, and Mo concentrations in the periprosthetic tissue (P = 0.0008, 0.029, and 0.007, respectively). Within the tissue adjacent to severely damaged implants, we measured median Co, Cr, and Mo concentrations of 7.81, 5.26, and 0.713 μg/mL, respectively. For minimally damaged implants, we report median Co, Cr, and Mo concentrations of 0.111, 1.80, and 0.179 μg/mL, respectively. In several of the 51 (14%) tissue samples, we measured Co and Cr concentrations > 10 μg/mL. Additionally, within periprosthetic tissues of devices with Ti tibial trays, Ti concentrations increased (P = 0.0052) arising, in part, from tibial-femoral component contact during arthroplasty. CONCLUSIONS: We (1) showed elevated periprosthetic tissue metal concentrations in TKA patients and (2) established a positive correlation between damage severity and subsequent metal release. Measured tissue metal concentrations approached the magnitudes reported following metal-on-metal THA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Bioengineering, Clemson University, Clemson, South Carolina; The Clemson University-Medical University of South Carolina Bioengineering Program, Charleston, South Carolina.
- Department of Biomedical Engineering, Drexel University, Implant Research Core, Philadelphia, Pennsylvania.
- Department of Orthopaedic Surgery, Cleveland Clinic Adult Reconstruction Research (CCARR), Cleveland Clinic, Cleveland, Ohio; Department of Biomedical Engineering, Cleveland Clinic, Cleveland, Ohio.
- Campbell Clinic Department of Orthopaedic Surgery and Biomedical Engineering, University of Tennessee Health Science Center, Memphis, Tennessee.
Zitieren
APA
Kurtz, P. W., Aslani, S., Kurtz, M. A., Taylor, L. M., Barnes, E. R., MacDonald, D. W., Piuzzi, N. S., Mihalko, W. M., Kurtz, S. M., & Gilbert, J. L. (2025). Cobalt-Chromium-Molybdenum Femoral Knee Implant Damage Correlates With Elevated Periprosthetic Metal Concentrations. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2025.02.075
Vancouver
Kurtz PW, Aslani S, Kurtz MA, Taylor LM, Barnes ER, MacDonald DW, et al. Cobalt-Chromium-Molybdenum Femoral Knee Implant Damage Correlates With Elevated Periprosthetic Metal Concentrations. The Journal of Arthroplasty. 2025. doi:10.1016/j.arth.2025.02.075
BibTeX
@article{Kurtz2025,
author = {Kurtz, Peter W and Aslani, Shabnam and Kurtz, Michael A and Taylor, Lilliana M and Barnes, Emma R and MacDonald, Daniel W and Piuzzi, Nicolas S and Mihalko, William M and Kurtz, Steven M and Gilbert, Jeremy L},
title = {Cobalt-Chromium-Molybdenum Femoral Knee Implant Damage Correlates With Elevated Periprosthetic Metal Concentrations},
journal = {The Journal of Arthroplasty},
year = {2025},
doi = {10.1016/j.arth.2025.02.075}
}
EUVIMED