EUVIMEDEUVIMED entdecken
ÜbersichtsarbeitOpen Access · CC BY 4.0

Standardization of extracellular vesicle concentration measurements by flow cytometry: the past, present, and future

Britta A Bettin, Zoltán Varga, Rienk Nieuwland, Edwin van der Pol

Journal of Thrombosis and Haemostasis · 2023-08 · 21 (8) · 2032-2044

DOI: 10.1016/j.jtha.2023.04.042 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Journal of Thrombosis and Haemostasis, 2023-08. Verfasst von Britta A Bettin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Experimental Clinical Chemistry, Amsterdam UMC location University of Amsterdam, Amsterdam, the Netherlands; Biomedical Engineering and…. Hauptthemen laut MEDLINE: Extracellular Vesicles. Die Arbeit wurde bisher 41-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Horizon 2020 Framework Programme, European Metrology Programme for Innovation and Research, Horizon 2020.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Concentrations of extracellular vesicles (EVs) in body fluids are being explored as disease biomarkers. Most laboratories use flow cytometry to characterize single EVs at high throughput. A flow cytometer (FCM) detects light scattering and fluorescence intensities of EVs. However, detection of EVs by flow cytometry is complicated for 2 reasons. First, EVs are small and have weak light scattering and fluorescence signals compared to cells and are, therefore, hard to detect. Second, FCMs differ in sensitivity and provide data in arbitrary units, which complicates data interpretation. Due to the mentioned challenges, the measured concentration of EVs by flow cytometry is cumbersome to compare between FCMs and institutes. To improve comparability, standardization and development of traceable reference materials to calibrate all aspects of an FCM are needed, as are interlaboratory comparison studies. Within this article, we will provide an overview of the standardization of EV concentration measurements, including the current effort to introduce robust calibration of FCMs, thereby enabling comparable concentration measurements of EVs, which in turn can be used to establish clinically relevant reference ranges of EV concentrations in blood plasma and other body fluids.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Extracellular VesiclesHumansFlow CytometryPlasmaCalibrationReference Standards

Stichwörter der Autor:innen

calibrationextracellular vesiclesflow cytometryinterlaboratory comparison studystandardization

Autor:innen und Institutionen

  • Laboratory of Experimental Clinical Chemistry, Amsterdam UMC location University of Amsterdam, Amsterdam, the Netherlands; Biomedical Engineering and Physics, Amsterdam, the Netherlands; Vesicle Observation Center, Amst…
  • Biological Nanochemistry Research Group, Institute of Materials and Environmental Chemistry, Research Centre for Natural Sciences, Budapest, Hungary.
  • Laboratory of Experimental Clinical Chemistry, Amsterdam UMC location University of Amsterdam, Amsterdam, the Netherlands; Biomedical Engineering and Physics, Amsterdam, the Netherlands.

Förderung

  • Horizon 2020 Framework Programme
  • European Metrology Programme for Innovation and Research
  • Horizon 2020
  • Magyar Tudományos Akadémia

Zitieren

APA

Bettin, B. A., Varga, Z., Nieuwland, R., & van der Pol, E. (2023). Standardization of extracellular vesicle concentration measurements by flow cytometry: the past, present, and future. Journal of Thrombosis and Haemostasis, 21(8), 2032-2044. https://doi.org/10.1016/j.jtha.2023.04.042

Vancouver

Bettin BA, Varga Z, Nieuwland R, van der Pol E. Standardization of extracellular vesicle concentration measurements by flow cytometry: the past, present, and future. Journal of Thrombosis and Haemostasis. 2023;21(8):2032-2044. doi:10.1016/j.jtha.2023.04.042

BibTeX
@article{Bettin2023,
  author = {Bettin, Britta A and Varga, Zoltán and Nieuwland, Rienk and van der Pol, Edwin},
  title = {Standardization of extracellular vesicle concentration measurements by flow cytometry: the past, present, and future},
  journal = {Journal of Thrombosis and Haemostasis},
  year = {2023},
  volume = {21},
  number = {8},
  pages = {2032--2044},
  doi = {10.1016/j.jtha.2023.04.042}
}
Kostenlos starten →