EUVIMEDEUVIMED entdecken
Crossref

Point-of-Care Bedside Brain Magnetic Resonance Imaging Is Safe in Extracorporeal Membrane Oxygenation Patients With Swan Ganz Catheters: A Phantom Experiment and Single Center Experience

Ifeanyi D Chinedozi, Eddy Boskamp, Zachary Darby, Jin Kook Kang, Hannah Rando, Haris Sair, John Pitt, Christopher Wilcox, Bo Soo Kim, Shivalika Khanduja, Glenn Whitman, Sung-Min Cho

Journal of Surgical Research · 2024

DOI: 10.1016/j.jss.2024.04.045 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Surgical Research, 2024. Verfasst von Ifeanyi D Chinedozi und 11 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiac Surgery, Johns Hopkins University, Baltimore, Maryland. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Extracorporeal Membrane Oxygenation, Phantoms, Imaging, Point-of-Care Systems. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 19 Literaturangaben. Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: More than 1.2 million pulmonary artery catheters (PACs) are used in cardiac patients per annum within the United States. However, it is contraindicated in traditional 1.5 and 3T magnetic resonance imaging (MRI) scans. We aimed to test preclinical and clinical safety of using this imaging modality given the potential utility of needing it in the clinical setting. METHODS: We conducted two phantom experiments to ensure that the electromagnetic field power deposition associated with bare and jacketed PACs was safe and within the acceptable limit established by the Food and Drug Administration. The primary end points were the safety and feasibility of performing Point-of-Care (POC) MRI without imaging-related adverse events. We performed a preclinical computational electromagnetic simulation and evaluated these findings in nine patients with PACs on veno-arterial extracorporeal membrane oxygenation. RESULTS: The phantom experiments showed that the baseline point specific absorption rate through the head averaged 0.4 W/kg. In both the bare and jacketed catheters, the highest net specific absorption rates were at the neck entry point and tip but were negligible and unlikely to cause any heat-related tissue or catheter damage. In nine patients (median age 66, interquartile range 42-72 y) with veno-arterial extracorporeal membrane oxygenation due to cardiogenic shock and PACs placed for close hemodynamic monitoring, POC MRI was safe and feasible with good diagnostic imaging quality. CONCLUSIONS: Adult ECMO patients with PACs can safely undergo point-of-care low-field (64 mT) brain MRI within a reasonable timeframe in an intensive care unit setting to assess for acute brain injury that might otherwise be missed with conventional head computed tomography.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingExtracorporeal Membrane OxygenationPhantoms, ImagingPoint-of-Care SystemsBrainCatheterization, Swan-GanzHumansMaleMiddle AgedFemaleAgedAdultFeasibility Studies

Stichwörter der Autor:innen

Acute brain injury (ABI)Cardiogenic shockExtracorporeal membrane oxygenation (ECMO)Hemodynamic monitoringPortable MRIPulmonary artery catheter (PAC)

Autor:innen und Institutionen

  • Division of Cardiac Surgery, Johns Hopkins University, Baltimore, Maryland.
  • Hyperfine Research, Guilford, Connecticut.
  • Department of Radiology, Johns Hopkins University, Baltimore, Maryland.
  • Critical Care Medicine, Mercy Hospital of Buffalo, Buffalo, New York.
  • Department of Neurology, Johns Hopkins University, Baltimore, Maryland. Electronic address: csungmi1@jhmi.edu.

Förderung

  • NHLBI NIH HHS
  • National Heart, Lung, and Blood Institute

Zitieren

APA

Chinedozi, I. D., Boskamp, E., Darby, Z., Kang, J. K., Rando, H., Sair, H., Pitt, J., Wilcox, C., Kim, B. S., Khanduja, S., Whitman, G., & Cho, S. M. (2024). Point-of-Care Bedside Brain Magnetic Resonance Imaging Is Safe in Extracorporeal Membrane Oxygenation Patients With Swan Ganz Catheters: A Phantom Experiment and Single Center Experience. Journal of Surgical Research. https://doi.org/10.1016/j.jss.2024.04.045

Vancouver

Chinedozi ID, Boskamp E, Darby Z, Kang JK, Rando H, Sair H, et al. Point-of-Care Bedside Brain Magnetic Resonance Imaging Is Safe in Extracorporeal Membrane Oxygenation Patients With Swan Ganz Catheters: A Phantom Experiment and Single Center Experience. Journal of Surgical Research. 2024. doi:10.1016/j.jss.2024.04.045

BibTeX
@article{Chinedozi2024,
  author = {Chinedozi, Ifeanyi D and Boskamp, Eddy and Darby, Zachary and Kang, Jin Kook and Rando, Hannah and Sair, Haris and Pitt, John and Wilcox, Christopher and Kim, Bo Soo and Khanduja, Shivalika and Whitman, Glenn and Cho, Sung-Min},
  title = {Point-of-Care Bedside Brain Magnetic Resonance Imaging Is Safe in Extracorporeal Membrane Oxygenation Patients With Swan Ganz Catheters: A Phantom Experiment and Single Center Experience},
  journal = {Journal of Surgical Research},
  year = {2024},
  doi = {10.1016/j.jss.2024.04.045}
}
Kostenlos starten →