Polyethylene wear testing of a nonmechanically linked total elbow replacement
Journal of Shoulder and Elbow Surgery · 2025
DOI: 10.1016/j.jse.2024.07.015 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Shoulder and Elbow Surgery, 2025. Verfasst von Brody K Nelson und 4 weiteren Autor:innen. Beteiligte Einrichtung: Element Materials Technology, Fairfield, OH, USA. Hauptthemen laut MEDLINE: Polyethylene, Materials Testing, Arthroplasty, Replacement, Elbow, Elbow Prosthesis. Die Arbeit nennt 53 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: The objective of this study was to perform a polyethylene wear test on a nonmechanically linked total elbow arthroplasty implant using a clinically relevant in-vitro elbow wear test methodology that simulated 10 years of use in the light to moderate activity of daily living range. MATERIALS AND METHODS: The test protocol applied an 80° arc of ulnohumeral motion beginning at 30° shy of full extension and progressing to 110° of flexion. Force was applied at 7° to recreate a valgus load on the elbow. A variable joint load profile at a frequency of 0.5 Hz was employed. The implants were tested for 5 million cycles (mc) in a bovine serum lubricant. Implant component failure was characterized and polyethylene wear was determined gravimetrically. RESULTS: After 5 mc, the small polyethylene bushing wear rate was 0.56 mg/mc. The medium size wear rate was 0.28 mg/mc. Three large sizes were tested and the average wear rate was 0.39 ± 0.07 mg/mc. No implant failure was identified. CONCLUSION: The test recreated an in vivo loading environment and measured polyethylene wear rates at specified cycle counts. The test demonstrated less wear than other joint replacements. Further clinical evaluation is necessary to determine if this translates into reduced complications of total elbow replacement associated with wear.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Element Materials Technology, Fairfield, OH, USA.
- Department of Orthopedic Surgery, University of Pittsburgh Medical Center, Pittsburgh PA, USA.
- Design Engineer, Arrch Orthopedics, Pittsburgh, PA, USA.
- Department of Orthopedic Surgery, University of Pittsburgh Medical Center, Pittsburgh PA, USA. Electronic address: Kaufra@upmc.edu.
Zitieren
APA
Nelson, B. K., Nazzal, E. M., Moody, D., Jakucki, M., & Kaufmann, R. A. (2025). Polyethylene wear testing of a nonmechanically linked total elbow replacement. Journal of Shoulder and Elbow Surgery. https://doi.org/10.1016/j.jse.2024.07.015
Vancouver
Nelson BK, Nazzal EM, Moody D, Jakucki M, Kaufmann RA. Polyethylene wear testing of a nonmechanically linked total elbow replacement. Journal of Shoulder and Elbow Surgery. 2025. doi:10.1016/j.jse.2024.07.015
BibTeX
@article{Nelson2025,
author = {Nelson, Brody K and Nazzal, Ehab M and Moody, Devon and Jakucki, Maciej and Kaufmann, Robert A},
title = {Polyethylene wear testing of a nonmechanically linked total elbow replacement},
journal = {Journal of Shoulder and Elbow Surgery},
year = {2025},
doi = {10.1016/j.jse.2024.07.015}
}
EUVIMED