New generation of superior single plating vs. low-profile dual minifragment plating in diaphyseal clavicle fractures: a biomechanical comparative study

Tatjana Pastor, Ivan Zderic, Till Berk, Firas Souleiman, Esther Vögelin, Frank J P Beeres, Boyko Gueorguiev, Torsten Pastor

Journal of Shoulder and Elbow Surgery · 2024-02 · 33 (2) · 409-416

DOI: 10.1016/j.jse.2023.08.008 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Shoulder and Elbow Surgery, 2024-02. Verfasst von Tatjana Pastor und 7 weiteren Autor:innen. Beteiligte Einrichtung: AO Research Institute Davos, Davos, Switzerland; Department for Plastic and Hand Surgery, Inselspital University Hospital Bern, University of Bern, B…. Hauptthemen laut MEDLINE: Clavicle, Fractures, Bone. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: SUVA.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Recently, a new generation of superior clavicle plates was developed featuring the variable-angle locking technology for enhanced screw positioning and a less prominent and optimized plate-to-bone fit design. On the other hand, minifragment plates in dual plating mode have demonstrated promising clinical results. The aim of the current study was to compare the biomechanical competence of single superior plating using the new-generation plate vs. dual plating using low-profile minifragment plates. METHODS: Sixteen paired human cadaveric clavicles were pairwise assigned to 2 groups for instrumentation with either a superior 2.7-mm variable-angle locking compression plate (group 1), or with one 2.5-mm anterior combined with one 2.0-mm superior matrix mandible plate (group 2). An unstable clavicle shaft fracture (AO/OTA 15.2C) was simulated by means of a 5-mm osteotomy gap. Specimens were cyclically tested to failure under craniocaudal cantilever bending, superimposed with bidirectional torsion around the shaft axis, and monitored via motion tracking. RESULTS: Initial construct stiffness was significantly higher in group 2 (9.28 ± 4.40 N/mm) compared to group 1 (3.68 ± 1.08 N/mm), P = .003. The amplitudes of interfragmentary motions in terms of axial and shear displacement, fracture gap opening and torsion, over the course of 12,500 cycles were significantly higher in group 1 compared to group 2, P ≤ .038. Cycles to 2 mm shear displacement were significantly lower in group 1 (22,792 ± 4346) compared to group 2 (27,437 ± 1877), P = .047. CONCLUSION: From a biomechanical perspective, low-profile 2.5/2.0-mm dual plates could be considered as a useful alternative for diaphyseal clavicle fracture fixation, especially in less common unstable fracture configurations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ClavicleFractures, BoneHumansBiomechanical PhenomenaFracture Fixation, InternalBone ScrewsBone Plates

Stichwörter der Autor:innen

Midshaft/diaphyseal clavicle fracturebiomechanicsdual platingimplant removalminifragment platesmotion trackingsymptomatic implants

Autor:innen und Institutionen

  • AO Research Institute Davos, Davos, Switzerland; Department for Plastic and Hand Surgery, Inselspital University Hospital Bern, University of Bern, Bern, Switzerland.
  • AO Research Institute Davos, Davos, Switzerland.
  • Department of Trauma, University Hospital Zurich, Zurich, Switzerland.
  • Department of Orthopaedics, Trauma and Plastic Surgery, University Hospital Leipzig, Leipzig, Germany.
  • Department for Plastic and Hand Surgery, Inselspital University Hospital Bern, University of Bern, Bern, Switzerland.
  • Department of Orthopaedic and Trauma Surgery, Lucerne Cantonal Hospital, Lucerne, Switzerland.

Förderung

  • SUVA

Zitieren

APA

Pastor, T., Zderic, I., Berk, T., Souleiman, F., Vögelin, E., Beeres, F. J. P., Gueorguiev, B., & Pastor, T. (2024). New generation of superior single plating vs. low-profile dual minifragment plating in diaphyseal clavicle fractures: a biomechanical comparative study. Journal of Shoulder and Elbow Surgery, 33(2), 409-416. https://doi.org/10.1016/j.jse.2023.08.008

Vancouver

Pastor T, Zderic I, Berk T, Souleiman F, Vögelin E, Beeres FJP, et al. New generation of superior single plating vs. low-profile dual minifragment plating in diaphyseal clavicle fractures: a biomechanical comparative study. Journal of Shoulder and Elbow Surgery. 2024;33(2):409-416. doi:10.1016/j.jse.2023.08.008

BibTeX
@article{Pastor2024,
  author = {Pastor, Tatjana and Zderic, Ivan and Berk, Till and Souleiman, Firas and Vögelin, Esther and Beeres, Frank J P and Gueorguiev, Boyko and Pastor, Torsten},
  title = {New generation of superior single plating vs. low-profile dual minifragment plating in diaphyseal clavicle fractures: a biomechanical comparative study},
  journal = {Journal of Shoulder and Elbow Surgery},
  year = {2024},
  volume = {33},
  number = {2},
  pages = {409--416},
  doi = {10.1016/j.jse.2023.08.008}
}