An innovative self-stabilised 3D-printed artificial vertebral body designed for clinical application and comparison with the conventional implants

Teng Zhang, Jian Li, Xinyu Li, Xin Pan, Xianlei Gao, Xiaojie Yang, Xiaolin Ma, Hao Li, Shiqing Feng, Zhongjun Liu

Journal of Orthopaedic Translation · 2025-07 · 53 · 52-62

DOI: 10.1016/j.jot.2025.04.010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Translation, 2025-07. Verfasst von Teng Zhang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, Qilu Hospital of Shandong University, Shandong University, Jinan, 250012, China. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Shandong Province Natural Science Foundation, Taishan Scholar Project of Shandong Province, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Corpectomy and bone defects reconstruction is a key surgical technique in spinal diseases treatment. Popular bone defect reconstruction methods include titanium mesh cage (TMC) fixation plate systems and traditional 3D-printed artificial vertebral body (3D-AVB). In our previous study, we conceptualised and created a self-stabilised 3D-printed artificial vertebral body (3D-SAVB) system and tested its clinical safety and efficacy, but have not compared with the conventional implants. This study was designed to compare our innovative 3D-SAVB system with a conventional 3D-AVB system, both mechanically and clinically.

Methods

This study included 33 patients with cervical spondylotic myelopathy who underwent single-level ACCF using the TMC, 3D-AVB, and 3D-SAVB systems. The operation time, intraoperative blood loss, neurological function recovery rate, average subsidence length, and cervical lordosis correction (C2-7 Cobb angle change) rates of the TMC, 3D-AVB, and 3D-SAVB groups were tested to compare their performance, and we selected four representative clinical cases with various diseases who underwent 3D-SAVB surgery for follow-up studies to demonstrate the clinical effect. In addition, finite element analysis was used to compare the stability, stress distribution, and artificial vertebral body stress of the 3D-SAVB, 3D-AVB, and TMC systems.

Results

The neurological function recovery rates of the TMC (84.8 ± 10.7 %), 3D-AVB (74.3 ± 7.9 %), and SAVB (85.99 ± 13.2 %) groups showed no significant difference (p > 0.05). The mean operation time of the TMC group (119.3 ± 21.5 min) is significant more than the 3D-SAVB (76.1 ± 23.1 mm) and 3D-AVB (82.6 ± 21.3 mm) groups (p < 0.05). The intraoperative blood loss of the TMC group (218.2 ± 51.5 ml) was significantly greater than that of the 3D-SAVB (187.6 ± 43.2 ml) and 3D-AVB groups (195.6 ± 31.3 ml) (p < 0.05). The mean subsidence length of the TMC group (3.5 ± 0.6 mm) was significantly greater than the 3D-AVB (1.3 ± 0.5 mm, p < 0.001) and 3D-SAVB (1.2 ± 1.1 mm, p = 0.002). The lordosis correction (C2-7 Cobb angle change) rate of the 3D-SAVB [(60.38 ± 6.2)%] and 3D-AVB [(61.4 ± 7.9)%] groups was significantly higher than that of the TMC group [(32.35 ± 3.7)%] (p < 0.05). Patients treated with the 3D-SAVB system achieved satisfactory treatment results with no postoperative complications during the follow-up period. The failed TMC fixation plate system underwent revision surgery using 3D-SAVB and demonstrated a superior prognosis. The biomechanical test showed that the 3D-SAVB system had greater longitudinal stability (p < 0.01), better stress distribution (p < 0.01), and less vertebral stress (p < 0.01) than the 3D-AVB and TMC systems.

Conclusion

These results demonstrate the mechanical advantages and great clinical application potential of our innovative 3D-SAVB system for challenging spinal surgeries. Translational Potential of this Article: The 3D-SAVB system eliminates the need for supplementary fixation plates, reduces surgical complications like subsidence, and improves cervical lordosis correction. Its biomechanical superiority and clinical efficacy support its adoption as a next-generation spinal implant.

Schlagwörter

Stichwörter der Autor:innen

Cervical lordosisSelf-stabilised 3D-printed artificial vertebral bodyStabilityStress distributionSubsidence

Autor:innen und Institutionen

  • Department of Orthopedics, Qilu Hospital of Shandong University, Shandong University, Jinan, 250012, China.
  • Beijing AK Medical Co., Ltd., China.
  • Department of Orthopedics, The Second Hospital of Shandong University, Jinan, 250033, China.
  • Department of Orthopedics, Peking University Third Hospital, Beijing, 100191, China.

Förderung

  • Shandong Province Natural Science Foundation
  • Taishan Scholar Project of Shandong Province
  • National Natural Science Foundation of China

Zitieren

APA

Zhang, T., Li, J., Li, X., Pan, X., Gao, X., Yang, X., Ma, X., Li, H., Feng, S., & Liu, Z. (2025). An innovative self-stabilised 3D-printed artificial vertebral body designed for clinical application and comparison with the conventional implants. Journal of Orthopaedic Translation, 53, 52-62. https://doi.org/10.1016/j.jot.2025.04.010

Vancouver

Zhang T, Li J, Li X, Pan X, Gao X, Yang X, et al. An innovative self-stabilised 3D-printed artificial vertebral body designed for clinical application and comparison with the conventional implants. Journal of Orthopaedic Translation. 2025;53:52-62. doi:10.1016/j.jot.2025.04.010

BibTeX
@article{Zhang2025,
  author = {Zhang, Teng and Li, Jian and Li, Xinyu and Pan, Xin and Gao, Xianlei and Yang, Xiaojie and Ma, Xiaolin and Li, Hao and Feng, Shiqing and Liu, Zhongjun},
  title = {An innovative self-stabilised 3D-printed artificial vertebral body designed for clinical application and comparison with the conventional implants},
  journal = {Journal of Orthopaedic Translation},
  year = {2025},
  volume = {53},
  pages = {52--62},
  doi = {10.1016/j.jot.2025.04.010}
}
Kostenlos starten →