Safety of robotic-assisted screw placement for spine surgery: Experience from the initial 125 cases

Tsutomu Akazawa, Yoshiaki Torii, Jun Ueno, Tasuku Umehara, Masahiro Iinuma, Atsuhiro Yoshida, Ken Tomochika, Seiji Ohtori, Hisateru Niki

Journal of Orthopaedic Science · 2024-07 · 29 (4) · 933-938

DOI: 10.1016/j.jos.2023.06.003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Science, 2024-07. Verfasst von Tsutomu Akazawa und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University School of Medic…. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Pedicle Screws. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The present study aimed to evaluate the safety of robot-assisted screw placement in 125 cases after introducing a spinal robotics system and to identify the situations where deviation was likely to occur. METHODS: The subjects were 125 consecutive patients who underwent robotic-assisted screw placement using a spinal robotics system (Mazor X Stealth Edition, Medtronic) from April 2021 to January 2023. The 1048 screws placed with robotic assistance were evaluated. We investigated intraoperative adverse events of the robotics system and complications occurring within 30 days after surgery. We evaluated screw accuracy and deviation and compared them for vertebral levels, screw insertion methods (open traditional pedicle screw [Open-PS], cortical bone trajectory screw [CBT], percutaneous pedicle screw [PPS], and S2 alar iliac screw [S2AIS]), diagnosis, and phases of surgical cases. RESULTS: The deviation rate of robotic-assisted screw placement for spine surgery was 2.2%. Complications were reoperation due to implant-related neurological deficit in 0.8% and surgical site infection in 0.8%. There was significant difference in the deviation rate between vertebral levels. The deviation rate of the T1-T4 level was high at 10.0%. There was significant difference in the deviation rate between Open-PS, CBT, PPS, and S2AIS. The PPSs had a high deviation rate of 10.3%. The deviation rates were not significantly different between patients with and without deformity. The deviation rate did not change depending on the experience of surgical cases, and the deviation rate was favorable from the onset. CONCLUSION: Although the robotic-assisted screw placement was safe, we should be extra vigilant when placing screws in the upper thoracic region (deviation rate 10.0%) and when using PPSs (deviation rate 10.3%).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresPedicle ScrewsHumansMaleFemaleMiddle AgedAgedAdultSpinal FusionPostoperative ComplicationsRetrospective StudiesYoung AdultAged, 80 and overAdolescentReoperation

Stichwörter der Autor:innen

Accuracy rateComplicationDeviation rateIntraoperative adverse eventRobotic-assisted pedicle screw placementRobotics

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University School of Medicine Hospital, Kawasaki, Japan. Electronic address: cds00350@par.odn.ne…
  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University School of Medicine Hospital, Kawasaki, Japan.
  • Department of Orthopaedic Surgery, Graduate School of Medicine, Chiba University, Chiba, Japan.
  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Akazawa, T., Torii, Y., Ueno, J., Umehara, T., Iinuma, M., Yoshida, A., Tomochika, K., Ohtori, S., & Niki, H. (2024). Safety of robotic-assisted screw placement for spine surgery: Experience from the initial 125 cases. Journal of Orthopaedic Science, 29(4), 933-938. https://doi.org/10.1016/j.jos.2023.06.003

Vancouver

Akazawa T, Torii Y, Ueno J, Umehara T, Iinuma M, Yoshida A, et al. Safety of robotic-assisted screw placement for spine surgery: Experience from the initial 125 cases. Journal of Orthopaedic Science. 2024;29(4):933-938. doi:10.1016/j.jos.2023.06.003

BibTeX
@article{Akazawa2024,
  author = {Akazawa, Tsutomu and Torii, Yoshiaki and Ueno, Jun and Umehara, Tasuku and Iinuma, Masahiro and Yoshida, Atsuhiro and Tomochika, Ken and Ohtori, Seiji and Niki, Hisateru},
  title = {Safety of robotic-assisted screw placement for spine surgery: Experience from the initial 125 cases},
  journal = {Journal of Orthopaedic Science},
  year = {2024},
  volume = {29},
  number = {4},
  pages = {933--938},
  doi = {10.1016/j.jos.2023.06.003}
}