EUVIMEDEUVIMED entdecken
Crossref

Accuracy of robotic-assisted pedicle screw placement comparing junior surgeons with expert surgeons: Can junior surgeons place pedicle screws as accurately as expert surgeons?

Yoshiaki Torii, Jun Ueno, Masahiro Iinuma, Atsuhiro Yoshida, Hisateru Niki, Tsutomu Akazawa

Journal of Orthopaedic Science · 2023-09 · 28 (5) · 961-965

DOI: 10.1016/j.jos.2022.06.012 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Science, 2023-09. Verfasst von Yoshiaki Torii und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University Hospital, Kawas…. Hauptthemen laut MEDLINE: Pedicle Screws, Robotic Surgical Procedures, Surgery, Computer-Assisted, Spinal Fusion. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 18 Literaturangaben. Gefördert von: Medtronic.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The purpose of this study was to verify whether a spine robotic system was useful for junior surgeons. METHODS: Twenty-seven patients underwent posterior spinal fusion with open surgery using a spine robotic system (Mazor X Stealth Edition, Medtronic Inc., Dublin, Ireland) from April to August 2021. Pedicle screw insertions were performed by five surgeons. The surgeon and insertion time were recorded for each pedicle screw. Two surgeons who are board-certified spine surgeons by the Japanese Society for Spine Surgery and Related Research (JSSR) were defined as the expert surgeon group. Three surgeons who were training to acquire qualifications as JSSR board certified spine surgeons were defined as the junior surgeon group. In postoperative CT images, the deviation of 255 pedicle screws was evaluated using the Gertzbein-Robbins (GR) grades. RESULTS: In the expert surgeon group, the GR grades were Grade A for 79 screws (90.8%), Grade B for 6 (6.9%), Grade C for 2 (2.3%), and 0 (0%) for Grades D and E. I In the junior surgeon group, the GR grades were Grade A for 162 screws (96.4%), Grade B for 6 (3.6%), and 0 (0%) for Grades C, D, and E. There was no significant difference in the deviation rate between surgeon groups (p = 0.08). The mean insertion times were 174.5 ± 83.0 s in the expert surgeon group and 191.0 ± 111.0 s in the junior surgeon group. There was no significant difference in the insertion time between surgeon groups (p = 0.22). CONCLUSIONS: There were no significant differences in the deviation rate and the insertion time of robotic-assisted pedicle screw placement between expert surgeons and junior surgeons who were training to acquire qualifications as JSSR board certified spine surgeons. Robotic-assisted pedicle screw placement can be effectively employed by junior surgeons.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pedicle ScrewsRobotic Surgical ProceduresSurgery, Computer-AssistedSpinal FusionSurgeonsHumansSpine

Stichwörter der Autor:innen

Deviation rate of pedicle screwExpert surgeonInsertion timeJunior surgeonRobotic-assisted pedicle screw placementRobotic-assisted spine surgerySpine robotic system

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University Hospital, Kawasaki, Japan.
  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan.
  • Department of Orthopaedic Surgery, St. Marianna University School of Medicine, Kawasaki, Japan; Spine Center, St. Marianna University Hospital, Kawasaki, Japan. Electronic address: cds00350@par.odn.ne.jp.

Förderung

  • Medtronic

Zitieren

APA

Torii, Y., Ueno, J., Iinuma, M., Yoshida, A., Niki, H., & Akazawa, T. (2023). Accuracy of robotic-assisted pedicle screw placement comparing junior surgeons with expert surgeons: Can junior surgeons place pedicle screws as accurately as expert surgeons?. Journal of Orthopaedic Science, 28(5), 961-965. https://doi.org/10.1016/j.jos.2022.06.012

Vancouver

Torii Y, Ueno J, Iinuma M, Yoshida A, Niki H, Akazawa T. Accuracy of robotic-assisted pedicle screw placement comparing junior surgeons with expert surgeons: Can junior surgeons place pedicle screws as accurately as expert surgeons?. Journal of Orthopaedic Science. 2023;28(5):961-965. doi:10.1016/j.jos.2022.06.012

BibTeX
@article{Torii2023,
  author = {Torii, Yoshiaki and Ueno, Jun and Iinuma, Masahiro and Yoshida, Atsuhiro and Niki, Hisateru and Akazawa, Tsutomu},
  title = {Accuracy of robotic-assisted pedicle screw placement comparing junior surgeons with expert surgeons: Can junior surgeons place pedicle screws as accurately as expert surgeons?},
  journal = {Journal of Orthopaedic Science},
  year = {2023},
  volume = {28},
  number = {5},
  pages = {961--965},
  doi = {10.1016/j.jos.2022.06.012}
}