Development and validation of a highly-sensitive, quantitative LC-MS/MS assay to evaluate plasma oxytocin
Journal of Mass Spectrometry and Advances in the Clinical Lab · 2025-04 · 36 · 19-28
DOI: 10.1016/j.jmsacl.2025.02.002 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Mass Spectrometry and Advances in the Clinical Lab, 2025-04. Verfasst von E Grifnée und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Chemistry, University Hospital of Liège, Belgium. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Oxytocin is a 9-amino acid peptide that serves as neuromodulator in the human central nervous system. This peptide is implicated in the regulation of diverse behaviors and plays a significant role in positive social interaction. Currently, oxytocin levels are measured using immunoassays. However, these methods have several limitations that can lead to false results and erroneous interpretation. Given the remarkably low endogenous level of oxytocin in human plasma (low ng/L levels), we developed and rigorously validated a novel and highly sensitive LC-MS/MS method for oxytocin quantification in plasma.
Methods
Oxytocin was initially extracted using solid-phase extraction with an Oasis HLB 30 mg plate and then subjected to LC-MS/MS analysis. PBS-0.1 % BSA served as surrogate matrix for the preparation of validation samples and the calibration curve, ensuring no endogenous interference. The validation design followed the Clinical Laboratory Standards Institute guidelines. Precision, accuracy, and measurement uncertainty were determined using single-nested analysis of variance and e.noval software.
Results
A lower limit of quantification of 1 ng/L was achieved. The method was validated for oxytocin concentrations ranging from 1 ng/L to 75 ng/L, with precision (coefficient of variation) below 10 %, accuracy ranging from 94 % to 108 %, and measurement uncertainty below 15 %.
Conclusion
In this work, we developed and validated a highly sensitive LC-MS/MS method for the quantification of oxytocin in plasma. Our novel methodology is well-suited for clinical applications.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Chemistry, University Hospital of Liège, Belgium.
- Department of Clinical Chemistry, CIRM, University of Liège, Belgium.
- Department of Psychiatry, University Hospital of Liège, Belgium.
Zitieren
APA
Grifnée, E., Mackowiak, A., Demeuse, J., Schoumacher, M., Huyghebaert, L., Determe, W., Dubrowski, T., Massonnet, P., Peeters, S., Scantamburlo, G., Cavalier, E., & Goff, C. L. (2025). Development and validation of a highly-sensitive, quantitative LC-MS/MS assay to evaluate plasma oxytocin. Journal of Mass Spectrometry and Advances in the Clinical Lab, 36, 19-28. https://doi.org/10.1016/j.jmsacl.2025.02.002
Vancouver
Grifnée E, Mackowiak A, Demeuse J, Schoumacher M, Huyghebaert L, Determe W, et al. Development and validation of a highly-sensitive, quantitative LC-MS/MS assay to evaluate plasma oxytocin. Journal of Mass Spectrometry and Advances in the Clinical Lab. 2025;36:19-28. doi:10.1016/j.jmsacl.2025.02.002
BibTeX
@article{Grifne2025,
author = {Grifnée, E and Mackowiak, A and Demeuse, J and Schoumacher, M and Huyghebaert, L and Determe, W and Dubrowski, T and Massonnet, P and Peeters, S and Scantamburlo, G and Cavalier, E and Goff, C Le},
title = {Development and validation of a highly-sensitive, quantitative LC-MS/MS assay to evaluate plasma oxytocin},
journal = {Journal of Mass Spectrometry and Advances in the Clinical Lab},
year = {2025},
volume = {36},
pages = {19--28},
doi = {10.1016/j.jmsacl.2025.02.002}
}
EUVIMED